. Within the proximity of equipment that supplies nitrogen gas. (flow rate : more than 3
λ
/min).
. Nitrogen gas tube : 3 m in length, inside diameter : 9.5 mm.
2.3 Reassembly
Note
: Reassembly of the instrument is performed by your local JASCO distributor.
2.3.1 Removing the cushion from the main unit
CAUTION
: When removing the cushion, be careful not to give an impact to the cam.
The cam and lever on the bottom of the instrument are separated by a cushion and
secured with a rubber band to protect the cam from damage and the wavelength from
shifting during transit.
Remove the cushion and rubber band after installing the main unit.
(1) Remove the side cover from the main unit.
Side cover
Setscrews
Fig. 2.1 Removing the side cover
(2) Remove the cushion and rubber band.
Cushion
Wavelength
lever
Rubber band
Cam
Fig. 2.2 Removing the cushion and rubber band