GO-5000M-PGE / GO-5000C-PGE
- 36 -
Note: This figure is if the trigger is set to ON. When the trigger is OFF,
FrameActive becomes True at the same timing of AcquisitionActive.
Fig.24
Multi Frame operation
◆
Forcing acquisition to stop
While AcquisitionActive is “TRUE”, if AcquisitionStop or AcquisitionAbort is
initiated, AcquisitionActive becomes “FALSE” (stop capturing).
Once the operation is set to “FALSE”, the internal FrameCount is reset.
However, if AcqusitionStop command is initiated during image output period,
AcqusitionActive becomes “FALSE” (stop capturing) after image output is completed.
Once, AcqusitionActive becomes “FALSE”, the internal count is reset.
Acqusition Frame Count : Can be set in the range of 1 to 255
8.1.1.3 Continuous
In this mode, when the AcquisitionStart command is set, the image is continuously output at the
current frame rate. This is the default setting for the GO-5000M-PGE and GO-5000C-PGE.
◆
Normal continuous operation
1) AcquisitionStart command is input
2) AcquisitionTriggerWait becomes effective
3) AcquisitionActive becomes
“
TRUE”
4) Images begin outputting continuously
5) AcquisitionStop command is sent
6) AcquisitionActive becomes
“
FALSE”. At this moment, the output stops.
However, if AcqusitionStop command is initiated during image output period,
AcqusitionActive becomes “FALSE” (stop capturing) after image output is completed.