3) Wait till “
JADEVER SCALE CO., LTD
” appears and flashes on weight window, put weights of 1/3 of full load on
(e.g. 15k model, 1/3 of full load is 5kg.)
4) Wait till “
” appears and flashes on total weight screen, put weights of 2/3 of full load
on (for 15k model, 2/3 of full load is 10kg)
5) Wait till
“
” appears and flashes on total weight screen, put weights of full load on
(for 15k model, full load is 15kg)
6) The calibration Procedure is finished with a symbol of “
away the weights.
7) Press key
TARE
to return to weighing mode.
4-4 Function Setting
” flashing, and then take
1) Turn on the power while pressing key
ZERO
and setting mode starts.
Note: The following Steps (2) ~ (11) do not require to operate in order.
2) Press numeric key
0
to shift backlight modes. Options are On, OFF and OnOFF.
=Auto-on with items greater than 9 divisions placed on the pan.
= No backlight
= Backlight
3) Press numeric key
2
t
o set the level in which the stable indication turns on (filtering).
The lower the setting, the faster stabilization time.
4) Press numeric key
3
to set the period of inactivity before the scale automatically turns
off.
Options are OFF(Non power-off)
、
5
、
10
、
30 and 60 (minutes).
5
) Press numeric key
4
to set the range in which the zero indication turns on .Options are
d0, d1, d2, d3, d4 and d5. (d= scale division)
6) Press numeric key
5
to set serial transmission rates. Options are 9600, 4800 and 2400.
If the selected external device is printer, please set the transmission rates as 9600.
13
JCL Service Manual