Chapter 7
– File Settings
84
5. Touch
Back
to complete the process.
H
OW TO
S
ET
A
UTO
S
AVE ON
S
HUTDOWN
When
“Save unsaved measurements to SD card on shutdown”
is enabled the instrument will prompt the
user to save all measurements in RAM to the SD Card before shutting down. The user will have 5 minutes (300
sec) to select one of the following options.
1. To save measurements on SD card, touch
Yes.
2. To shut down without saving, touch
No.
3. To go back to the previous menu and not shut down, touch
Cancel.
This prompt will not appear if
Auto Save
is enabled, as the measurements are automatically stored following a
measurement.
FIGURE 70 - AUTO SAVE ON SHUTDOWN PROMPT.
If no decision is made within 300 seconds (5 min) the instrument will power down and all current measurements
will be lost.