Chapter 5
– Setup Menu
70
→
D
EFINING
P
ASS
/
F
AIL
R
EGIONS
R
ECTANGLES
(
OR
S
QUARES
)
To define a Rectangle or square, two sets of values are entered. The first set is the CIE 1931 x, y coordinate of
the upper left hand corner referred to in the Pass / Fail menu as the rectangles are oriented so that the sides are
parallel to boundaries (axes) of the CIE diagram.
1. Touch the field adjacent to
CIE Pass / Fail
so that it reads
Enabled.
2. Touch the
▼▲ icons in the
Shape
field until
Rectangle
appears.
3. Define the Upper Left Corner by:
a.
Touching the ▼▲ icons in the
Edit
field until
Corner
appears.
b. Touch the gray field next to
Left
to display the data entry field for the
Upper
Left
CIE 1931 x
coordinate.
c.
Enter the value (including the decimal point) then touch
Back.
This value is the
absolute
position
of the x coordinate in the CIE 1931 diagram.
d. Touch the gray field next to
Top
to display the data entry field for the
Upper
Left CIE 1931 y
coordinate.
e. Enter the value (including the decimal point) then touch
Back.
This value is the
absolute
position
of the y coordinate in the CIE 1931 diagram.
4. Define the
Size
by:
a.
Touching the ▼▲ icons in the
Edit
field until
Size
appears.
b. Touch the gray field next to
Width
to display the data entry field for the
Lower Right
CIE 1931 x
coordinate.
c. Enter the value (including the decimal point) then touch
Back.
This value is the
distance
from
the Top Left x coordinate in Step 3c.
d. Touch the gray field next to
Top
to display the data entry field for the
Lower Right CIE 1931 y
coordinate.
e. Enter the value (including the decimal point) then touch
Back.
This value is the
distance
from
the Top Left y coordinate in Step 3e.
Note: To define a square, the Width and Height values must be identical. For a horizontal rectangle, the
Width value must be larger than the Height value, and for a Vertical rectangle, the Height value
must be larger than the Width value.
5. Following the measurement, touch the
< >
icons at the bottom of the result screen to scroll to the CIE
diagram screen.
Pass
or
Fail
appears in the lower right hand corner of the screen.
Содержание Photo Research SpectraScan PR-7 Series
Страница 2: ......
Страница 10: ...PR 7XX User s Manual vi...
Страница 19: ...Chapter 2 System Overview 15 To traverse back to the System Menu simply touch the Back button twice...
Страница 32: ......
Страница 84: ......
Страница 91: ...Chapter 7 File Settings 87 FIGURE 73 ILLUMINANCE LUMINANCE CONCEPT...
Страница 92: ......
Страница 102: ......
Страница 108: ......
Страница 117: ...Chapter 12 Connectivity 113...
Страница 118: ......