![Jadak Photo Research SpectraScan PR-7 Series Скачать руководство пользователя страница 123](http://html1.mh-extra.com/html/jadak/photo-research-spectrascan-pr-7-series/photo-research-spectrascan-pr-7-series_user-manual_3186341123.webp)
Appendix 1
– Remote Control Mode
119
R
EMOTE
C
ONTROL
C
OMMAND
D
ETAIL
Command
Description
A
Purpose
: Abort measurement in progress
Syntax
: A[CR]
Response
: 0000, [CR][LF] Measurement aborted.
0001, [CR][LF] No measurement in progress
B
Purpose
: Set LCD backlight level
Syntax
: Bnn[CR]
Bnn = Backlight / Brightness level in percentage.
Range of nn = 0 to 100%
Response
: Backlight set to nn %
C
Purpose
: Clears the current instrument error
Syntax
: C[CR]
Response
: None
D
Purpose
: Download data from the PR-7xx
Syntax
: D<data code>[CR]
Response
: 0000,<data>[CR][LF] If all OK, else
NNNN[CR][LF] (NNNNN = Error code)
Note: <data> in response code refers to the specific measurement data set returned based on
the data code sent to the instrument. Refer to the Data Code section for details..
E
Purpose
: Full Duplex (Echo) ON / OFF
Syntax
: E[CR]
Response
: None
F
Purpose
: Measure frequency of light source
Syntax
: F[CR]
Response
: 0000,ff.ff Hertz (Period = nnnnn milliseconds) If all OK else
NNNN[CR][LF] (NNNN = Error code)
I
Purpose
: Return instrument status / error report
Syntax
: I[CR]
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
L
Purpose
: Assign measurement description
Syntax
: L<Character String with max length of 20 characters>[CR]
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: Entry remains valid for the duration of the current Remote Mode session or until a new L
command is issued. If L[CR] is issued with an empty string, the current description is
returned.
M
Purpose
: Make a Measurement with the PR-7xx
Syntax
: M<data code>[CR]
Response
: 0000,<data>[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: <data> in response code refers to the specific measurement data set returned based on
the data code sent to the instrument. Refer to the Data Code section for specific
information.
O
Initialize Data Logger Mode. Data Logger Mode acts like a printer, automatically sending
measurement result data over the USB interface following a measurement. Data types can
be defined by the user through switches.
Note:
Data Logging is supported on PR-7XX firmware versions 2.58 and above only.
Contact Photo Research for upgrade information.
Содержание Photo Research SpectraScan PR-7 Series
Страница 2: ......
Страница 10: ...PR 7XX User s Manual vi...
Страница 19: ...Chapter 2 System Overview 15 To traverse back to the System Menu simply touch the Back button twice...
Страница 32: ......
Страница 84: ......
Страница 91: ...Chapter 7 File Settings 87 FIGURE 73 ILLUMINANCE LUMINANCE CONCEPT...
Страница 92: ......
Страница 102: ......
Страница 108: ......
Страница 117: ...Chapter 12 Connectivity 113...
Страница 118: ......