Appendix 1
– Remote Control Mode
124
Setup
Command
Description
SN
Cycles to Average
Defines the number of measurements (cycles) to average when calculating photometric
and colorimetric values. The average of the spectra are used to calculate other values.
The range of cycles to average is 1 to 99. The default is 1.
Syntax
: SNaa[CR]
Where
: aa = Cycles to Average Range 1 to 99
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
SO
CIE Observer
Photometric and Colorimetric values can be calculated using either CIE 2
or 10
Standard
Observer data sets. Use this specifier to choose the CIE data set for calculations for the
next measurement. The default is 2
.
Syntax
: SOn[CR]
Where
: n = CIE Observer
2 = 2
10 = 10
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
SP
Primary Accessory
A Primary Accessory is one that replaces the standard objective lens (typically the MS-75)
during use and can be used in conjunction with an Add-on Accessory.
Syntax
: SPnn[CR]
Where
: nn = Accessory Code
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: Accessory Codes can be found by running report 116 (command D116). See the
Data Codes section for specific details.
Содержание Photo Research SpectraScan PR-7 Series
Страница 2: ......
Страница 10: ...PR 7XX User s Manual vi...
Страница 19: ...Chapter 2 System Overview 15 To traverse back to the System Menu simply touch the Back button twice...
Страница 32: ......
Страница 84: ......
Страница 91: ...Chapter 7 File Settings 87 FIGURE 73 ILLUMINANCE LUMINANCE CONCEPT...
Страница 92: ......
Страница 102: ......
Страница 108: ......
Страница 117: ...Chapter 12 Connectivity 113...
Страница 118: ......