Chapter 3 Measurement Principle
17
3.2 Main Unit
As illustrated in Figure 3.2, the electromotive force e induced in Section 3.1
is brought in the input part of the main unit through the sensor cable.
As explained in Equation 3.1 in Section 3.1, the electromotive force e is
integrated by the integrator of the main unit, because it is the time
differentiation of the current.
The output voltage of the main unit Vout is expressed by Equation 3.2.
……
(3.2)
A:
Constant determined by the component constants of
the circuit of the main unit
The output voltage of the main unit (current waveform) is observed by an
oscilloscope and other monitoring devices through the BNC cable.
As expressed by the following Equation 3.3, the voltage component Vout
observed by monitoring devices can be converted to the current I with a
sensitivity S (mV/A) described on the label affixed to the back of the main
unit.
I = Vout
/
S
…… (3.3)
The current I (current waveform) of the conductor to be measured is
determined by the measurement principle described in this chapter.
Figure 3.2 Illustration of Rogowski Coil Operation
Main unit (in the dotted area)
Sensor cable (coaxial)
Input part
Integrator
Output part
BNC cable (coaxial)
Observation with an
oscilloscope (not included)
and other monitoring
d i
Содержание SS-663
Страница 1: ...Rogowski Coil Current Probe SS 663 664 665 Instruction Manual...
Страница 2: ...2018 IWATSU ELECTRIC CO LTD All rights reserved...
Страница 4: ...Memo...
Страница 16: ...Chapter 2 Names and Functions of Parts 12 Chapter 2 Names and Functions 2 1 Main Unit...
Страница 31: ...Chapter 5 Specifications 27 5 2 Appearance 1 Main Unit Unit mm...
Страница 32: ...Chapter 5 Specifications 28 2 Rogowski Coil Sensor Unit mm...
Страница 34: ...SS 663 664 665...