Chapter 4 Measurement
19
4.2 Measurement Procedure
(1) Make sure that all preparations described in the previous section are
completed.
(2) Set up a target to be measured so that it can be used.
Do not energize the target to be measured even when it is ready
to be used.
Warning
・
Do not energize a target to be measured before
attaching the Rogowski Coil of this instrument
to the target. It is dangerous to apply power to
the
target
to
be
measured
before
the
attachment.
(3) Place the target to be measured through the Rogowski Coil Sensor
part.
a) As illustrated in Figure 4.1, to release the lock of the Rogowski
Coil, be sure to turn the screw lock in a direction opposite the
direction indicated by “LOCK” until you feel a click. After releasing
the lock, remove the one end of the Rogowski Coil from the joint
hole.
b) Loop the Rogowski Coil around the conductor to be measured.
c) Insert the Rogowski Coil into the joint hole.
*: The insertion length of the coil is about 32 mm from the joint hole
end. Insert the coil until it clicks into place. To lock the Rogowski
Coil after the insertion, be sure to turn the screw lock in the
direction indicated by “LOCK” until you feeling a click. To achieve
accurate measurement, be sure to insert the end of the coil into
the joint hole securely.
Figure 4.1 How to Release Rogowski Coil
Joint hole
Rogowski Coil
Joint box
Screw lock
Содержание SS-293L
Страница 1: ...Rogowski Coil Current Probe SS 293S 293L 294S 294L SS 295S 295L 296S 296L Instruction Manual...
Страница 2: ...2015 IWATSU TEST INSTRUMENTS CORPORATION All rights reserved...
Страница 4: ...Memo...
Страница 16: ...Chapter 2 Names and Functions of Parts 12 Chapter 2 Names and Functions of Parts 2 1 Main Unit...
Страница 30: ...Chapter 5 Specifications 26 5 2 Appearance 1 Main Unit Unit mm...
Страница 31: ...Chapter 5 Specifications 27 2 Rogowski Coil Sensor Part Unit mm...
Страница 32: ...Chapter 5 Specifications 28 Memo...
Страница 34: ...SS 293S 293L 294S 294L 295S 295L 296S 296L...