Furnace head opens when this error occurs.
**
A program in progress is stopped.
***
The error cannot be acknowledged; the programs cannot be started.
34
Index Category Error
ERR
Conti-
Error Message Text
No.
nuation
possible
58
System
Final vacuum value not
800
The required final vacuum value cannot be reached. Check the vacuum pump.
reached
59
System
Vacuum drop
801
An unacceptable vacuum drop has occurred.
60
System
The vacuum does not
802
No vacuum increase could be measured. Check the following points: Is the firing chamber tight (no
increase (self-test)
contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?
Is the fuse F1 ok?
89
System
ATK2 calibration:
1302
Error during calibration. Sample may not be correctly inserted. Try again with a new sample and make sure the
Pre-heating to 962 °C
**
sample makes ample contact.
93
Note
Calibration reminder
1310
Some time has passed since the last calibration procedure. Calibrate the furance soon.
95
Note
Dehumidification reminder
1312
Some time has passed since the last dehumidification. Conduct a dehumidification in the near future.
100
Note
Temperature > VT at the
1510
The temperature in the firing chamber is higher than the pre-drying temperature. Press START to continue
start of a firing program
the program despite the error message.
101
Entry
Software update:
1520
The file selected for the software update is invalid. The update cannot be conducted.
Invalid file
102
Entry
Software update:
1521
The software to be installed is older or the same as the one installed on the furnace. The update cannot be
Invalid version
conducted.
103
System
Software update:
1522
An error has occurred during the software update. Do NOT switch off the furnace and try again.
Error during update
If the error reoccurs, try conducting the update via the USB interface.
8.2 Other error messages
Please contact the Ivoclar Vivadent After Sales Service, if one of the
following error messages is being displayed:
25, 29
43, 44, 45, 46, 47, 48
54, 56
103, 107
143, 144, 145, 146, 147, 148
700, 701, 703, 704, 705, 706, 707
1010, 1011, 1012, 1013, 1014, 1015, 1016, 1017, 1018, 1019
1024, 1025, 1026, 1028
1143, 1144, 1145, 1146, 1147, 1148
1202, 1203, 1204, 1205, 1206, 1207
1300, 1301, 1303, 1304, 1305
1400, 1401, 1402
1500, 1530, 1540
Содержание Programat P700/G2
Страница 1: ...Operating Instructions Programat P700 G2 Valid as of Software Version 5 0...
Страница 2: ...2...
Страница 5: ...40 38 34 6 10 56 35 20 51 55 23 56 33 36 52 22 7 2 1 3 9 5 8 5...
Страница 38: ...38 10 2 Menu structure 10 2 1 Possibilities of the program selection...
Страница 39: ...39 10 2 2 Overview of program groups 10 2 3 Settings information...