5.6 Description of the symbols in the display
Symbol Name
Pre-vacuum
One-stage
program
Two-stage
program
Standard opening of
the furnace head
Quick opening of
the furnace head
Cooling Shock
Protection
Crossed-out
Cooling Shock
Protection
Open lock
Closed lock
Crossed-out
"crescent moon"
"Crescent moon"
Individual group
write-protection
active
General write-
protection active
Last opened picture
Automatic vacuum
Crossed-out
Automatic vacuum
Open furnace head
with heat rays
Half open furnace
head with heat rays
Thermo Shock
Protection
Crossed-out
Thermo Shock
Protection
Meaning
Vacuum generation
starts before the
heating begins.
Abstract firing curve
of a one-stage
program
Abstract firing curve
of a two-stage
program
Furnace head opens
in the standard
period of time
Furnace head opens
in a short period of
time, i.e. quicker
(arrow)
Cooling Shock Pro-
tection is selected or
active
Cooling Shock Pro-
tection is inactive
Individual write-
protection inactive
Individual write-
protection active
Overnight program
function
Overnight program
function only active
for the next program
sequence
All 10 programs of
this group are write-
protected
All programs are
write-protected
By pressing the
symbol, the last
opened picture is
shown in the multi-
media area
Automatic vacuum
active
Automatic vacuum
inactive
Pre-drying with
completely open
furnace head active
Pre-drying with half
open furnace head
active
Thermo Shock
Protection active
Thermo Shock
Protection inactive
Symbol
5.5.5.2 Heater test
The quality of the heating muffle
may be automatically checked by
means of the heater test (dura-
tion: approximately 7 minutes).
The heater test should only be
conducted with the empty firing
chamber, since an object in the
chamber (e.g. firing tray) may
influence the test result. Conduct
the heater test immediately after
switching on the furnace and before any actual firing procedures. If
the furnace is too hot, an incorrect heating muffle quality will be
indicated. If the heating element quality falls below 50 %, replacing
the heating element is recommended.
5.5.5.3 Cleaning program
The cleaning program is used to "clean" the heating muffle
(duration: approx. 17 minutes). After a cleaning program, it is
recommended to calibrate the furnace. In case of problems with
discolouration of the ceramic, we recommend replacing the firing
table or the firing tray material.
5.5.5.4 Dehumidification program
The condensation of water in the insulation of the firing chamber
and the vacuum pump will result in a lower vacuum and thus to
impaired firing results. For that reason, the furnace head should be
kept closed when the furnace is switched off, in order to prevent
the absorption of humidity. Start the dehumidification program if
required (humidity in the insulation).
5.5.5.5 Keypad test
Each time the keypad is pressed, a short beep sounds. The keypad
test can be ended by pressing ESC.
5.5.5.6 Display test
(page 2/2)
Two different "chequer-board patterns" are alternately shown on the
entire display. This allows the visual check of each individual pixel.
The display test can be ended by pressing ESC.
Содержание Programat P700/G2
Страница 1: ...Operating Instructions Programat P700 G2 Valid as of Software Version 5 0...
Страница 2: ...2...
Страница 5: ...40 38 34 6 10 56 35 20 51 55 23 56 33 36 52 22 7 2 1 3 9 5 8 5...
Страница 38: ...38 10 2 Menu structure 10 2 1 Possibilities of the program selection...
Страница 39: ...39 10 2 2 Overview of program groups 10 2 3 Settings information...