Ivoclar Vivadent Programat P700/G2 Скачать руководство пользователя страница 29

29

6.4.12 Pre-vacuum

If a firing program with pre-
vacuum is conducted, the vacuum
pump is switched on at the end of
the closing time (as soon as the
furnace head is closed). Once the
pre-vacuum time has elapsed, the
heating phase begins. Upon the
start of a program with an
individually activated pre-vacuum
(value between 1:00 and 5:00),

the V1 value is ignored. The vacuum is maintained until V2 is
reached. V2 must be higher than the stand-by temperature B.

6.4.13 Overnight program

The overnight program function can be activated for the next
program sequence in the parameter list (touch button with the
"crescent moon" symbol).
If the overnight program function has been activated, the furnace
head remains open after the firing process, the heater is switched
off, and the green START LED flashes.
No melody is played if the temperature drops below a certain level.
Once the temperature has dropped below 100 °C / 212 °F, the
furnace head closes, the heater remains switched off, and the
furnace cools to room temperature.

Advantages of the overnight program function:
In case of a power failure, the overnight program is always
continued. The program resumes at the point at which the power
failure has occurred.

6.4.14 ”Quick cooling“

If the "Open furnace head" key is pressed again with the furnace
head already completely open, the "Quick cooling" function is
started. This means that the vacuum pump is switched on for
5 minutes. This function can be stopped early by pressing STOP,
"Close furnace head", or START. This function can be activated
anytime with the furnace head open.

6.4.15 Software Update

The user should be able to conduct a software update by CD, PC,
and download cable. For that purpose, the software download
mode of the furnace is activated by pressing two special keys
simultaneously while the power supply is switched on. For further
details, please refer to the Software Update Instructions
(www.ivoclarvivadent.com/downloadcenter).

A further method for updating the
software is to use the USB
memory stick. This item is
available in the Settings menu on
page 2.
To update the software, a USB
memory stick is required with a
current software image in the
form of a *.gz file. The software
version on the USB memory stick

must be newer than the one on the furnace (see Selection -
Information).

Step 1:

Connect the USB memory stick with the current software
version to the furnace

Step 2:

Select the menu item “Software Update”

Step 3:

Select the software image file (the file name has to start
with P700/G2_)

Step 4:

Start the software update with the “Open” key

Step 5:

Wait until the furnace indicates a successful update
process

Step 6:

Re-start the furnace

Important

Please note that the individual programs will be over-
written in the software update process. Therefore, we
recommend a data back-up (e.g. on a USB stick).

6.4.7 Help

Help text for the current screen.

6.4.8 Error message

The error group symbol should
supply a first indication of the type
of error (entry error = exclamation
mark; technical error = fork
wrench; Note = "i" symbol) with-
out the user having to consult the
Operating Instructions.

6.4.9 One-stage / two-stage programs

If the cursor is on the "one-stage program" symbol, pressing the
– / + key results in the symbol to change to the "two-stage
program" symbol. At the same time, the program was also changed
to become a "two-stage program".

If the cursor is on the "two-stage program" symbol, pressing the
– / + key results in the symbol to change to the "one-stage
program" symbol. At the same time, the program was also changed
to become a "one-stage program".

6.4.10 Program status indication

The current program status is indicated in the firing curve display:

pre-drying, closing, pre-vacuum, firing, long-term cooling, opening.

If the program is interrupted, "

Pause

" starts flashing as an indicator.

If a program is prematurely stopped, "

Vacuum release

" is flashing

while the firing chamber is flooded.

6.4.11 Pre-drying (one-stage or two-stage program)

Displaying the touch button "Pre-Drying Temperature":

Change to page 2/2 in the
"Settings" screen using the + key.
Then, select the touch button
"Preheating with the furnace head
open" and use the +/- keys to
switch on the "pre-heating"
setting.

In this way, the touch button "Pre-
drying temperature" is displayed
in the parameter list. The function
"pre-heating", however, is not yet
activated (temperature = 0).

For a program with individually
activated pre-drying, the desired
"pre-drying temperature" is set
after the program start with the
furnace head open (heating or
cooling). Once this temperature is
reached, pre-drying is carried out
during the "pre-drying holding
time" Once the holding time has
elapsed, the furnace head is
closed within the desired closing
time.

Содержание Programat P700/G2

Страница 1: ...Operating Instructions Programat P700 G2 Valid as of Software Version 5 0...

Страница 2: ...2...

Страница 3: ...djustable parameters and possible value ranges 5 5 Settings and information 5 6 Explanation of the symbols on the display 5 7 Explanation of the speaker signals 5 8 OSD Optical Status Display 5 9 Mult...

Страница 4: ...uple plug socket 32 Leaf spring 33 Air vents base 34 Cooling tray 35 Screw for cooling tray 36 Hood 37 Knurled screw for hood 38 Air vents furnace head 39 Air vents rear panel 40 Warnings 41 Furnace h...

Страница 5: ...40 38 34 6 10 56 35 20 51 55 23 56 33 36 52 22 7 2 1 3 9 5 8 5...

Страница 6: ...6 49 26 30 29 50 54 27 28 11 60 32 43 12 52 13 15 19 53 58 57 46 21 17 18 16 51 11 3 38 36 37 39 12 13 58 53 57 19...

Страница 7: ...wn 90 Open furnace head 91 Close furnace head 92 Numeric keys 93 Multimedia key 94 Home key 95 Help key 96 Power saving key 8 4 44 3 41 50 54 42 100 Programat firing tray 101 Metal pin A 102 Metal pin...

Страница 8: ...nts and have the following meanings Risks and dangers Important information Contraindication Burn hazard Risk of crushing The Operating Instructions must be read 1 3 Notes regarding the Operating Inst...

Страница 9: ...s for Use must be observed The furnace must be operated under the indicated environmental and operating conditions Chapter 9 The Programat P700 G2 must be properly maintained 2 1 1 Risks and dangers T...

Страница 10: ...d 10 2 1 11 Risk of crushing and burn hazard Never reach under the furnace head with the hand or other parts of the body during operation since there is a risk of crushing and a burn hazard 2 1 12 Con...

Страница 11: ...operation is 5 C to 40 C 41 F to 104 F If the furnace has been stored at very low temperatures or high atmospheric humidity the head has to be opened and the unit dried or left to adjust to room tempe...

Страница 12: ...pump The firing process is controlled with the corresponding electronic controls and software Moreover the set and actual temperatures are continuously compared The Programat P700 G2 has Ivoclar Vivad...

Страница 13: ...use the two corresponding inserts Fold the side flaps 2 and combine the two packaging parts by means of the transportation flaps The packaging may be disposed of with the regular household refuse 4 2...

Страница 14: ...r Lift the furnace head with both hands see picture and carefully position it on the furnace head mounting 43 Secure the cooling tray 34 with the two screws 35 including the silicone washer 47 Place t...

Страница 15: ...furnace base mounting mark 42 Make sure that the firing plate holder 48 is not damaged by mounting the furnace head Step 4 Connections Connect the cables of the furnace head with the furnace base Proc...

Страница 16: ...lishing additional connections Power connection Please make sure that the voltage indicated on the rating plate complies with the local power supply Connect the power cord 16 with the power socket 17...

Страница 17: ...hood 36 3 Disconnect the thermocouple plug 26 4 Disconnect the heater plug 28 5 Press the leaf spring 32 with a finger lift off the furnace head at the same time and remove it Make sure the furnace h...

Страница 18: ...set by means of the respective touch button After that the next basic setting screen temperature mode appears These settings are then saved and will no longer appear upon the subsequent start ups 4 5...

Страница 19: ...isplay shows the power saving icon Pressing any key ends the power saving function Key Function Program key Shows the currently selected program Pressing the key several times graphical representation...

Страница 20: ...ng the keys List of parameters Two stage program Automatic vacuum By activating the automatic vacuum in the parameter list the programming of a firing program with vacuum is supported For this purpose...

Страница 21: ...parameter field is activated by tapping the respective touch button and the settings can be edited using the keys e g Speaker Pressing the corresponding touch button activates the respective paramete...

Страница 22: ...to delete an individual character to the left of the cursor The entire description can be deleted by pressing the Delete all touch button 22 With Pictures background colour the display of pictures can...

Страница 23: ...an be edited using the keys or the numeric keypad If a memory stick is used an additional touch button is dis played This screen is displayed by pressing the touch button Copy The copy procedure is ex...

Страница 24: ...hermo Shock Protection inactive Symbol 5 5 5 2 Heater test The quality of the heating muffle may be automatically checked by means of the heater test dura tion approximately 7 minutes The heater test...

Страница 25: ...ultimedia menu OSD system 5 8 3 Deactivation In order to deactivate the OSD the brightness in the multimedia area see Chapter 5 9 Multimedia has to be set to 0 5 9 Multimedia Pressing the Multimedia k...

Страница 26: ...becomes audible If the player is stopped by means of the Stop button you can return to the mp3 file of the USB memory stick by pressing ESC Risks and dangers Listening to audio files with headphones a...

Страница 27: ...d the symbol is switched to the one stage symbol by pressing the or key the program has been set to one stage 6 1 3 Description of the firing curve display If the program is started with the START key...

Страница 28: ...ng program Press the STOP key once to pause a running program The green LED in the START key blinks Press the STOP key twice to completely stop the program or press START to continue 6 4 5 Changing th...

Страница 29: ...than the one on the furnace see Selection Information Step 1 Connect the USB memory stick with the current software version to the furnace Step 2 Select the menu item Software Update Step 3 Select th...

Страница 30: ...ce a prepared USB memory stick is ready for use the program selection screen will show another touch button for the selection of the UBS memory stick The active status of the USB memory stick is shown...

Страница 31: ...ance reasons subsequent temperature calibration is required monthly Vacuum hose Firing chamber Check if there is condensate in the vacuum hose or firing chamber daily Firing chamber Check the quartz g...

Страница 32: ...on the entire display This allows the visual check of each individual pixel The display test can be ended by pressing ESC 7 2 Cleaning The furnace may only be cleaned when it is cool since there is a...

Страница 33: ...may still be started 14 System Heating muffle defective 24 The condition of the muffle is so poor that is has to be replaced immediately 16 Entry T is B 160 C at the 26 Firing chamber too hot to star...

Страница 34: ...has passed since the last dehumidification Conduct a dehumidification in the near future 100 Note Temperature VT at the 1510 The temperature in the firing chamber is higher than the pre drying temper...

Страница 35: ...furnace off and on again Vacuum pump does not start working Is the vacuum pump fuse defective Check fuse and replace if necessary Was the maximum power consumption Use only the vacuum pump recommended...

Страница 36: ...305 mm 400 mm with cooling tray Height 320 mm Usable size of the firing chamber Diameter 80 mm Height 48 mm Max firing temperature 1200 C 2192 F Weight Furnace base 13 5 kg Furnace head 4 5 kg Safety...

Страница 37: ...nt Service Center Important information The current program table is also available at www ivoclarvivadent com downloadcenter The program tables can be downloaded from the Internet as PDF files Please...

Страница 38: ...38 10 2 Menu structure 10 2 1 Possibilities of the program selection...

Страница 39: ...39 10 2 2 Overview of program groups 10 2 3 Settings information...

Страница 40: ...Condesa 06170 M xico D F Mexico Tel 52 55 5062 1000 Fax 52 55 5062 1029 www ivoclarvivadent com mx Ivoclar Vivadent Ltd 12 Omega St Albany PO Box 5243 Wellesley St Auckland New Zealand Tel 64 9 914 9...