8
The use of this device close to other
equipment or stacked with it should be
avoided because the correct function can be
disrupted. If such use is unavoidable, the
devices need to be monitored and checked
for correct function.
Portable and mobile high-frequency
communication devices may interfere with
medical equipment. The use of mobile
phones during operation is not allowed.
Caution – The use of controls or adjustment
devices or performing procedures other than
those specified herein may result in
hazardous radiation exposure.
Warning
This unit should not be used near flammable
anaesthetics or mixtures of flammable
anaesthetics with air, oxygen or nitric oxide.
Usage and liability
– Bluephase Style 20i must only be employed for the
intended use. Any other uses are contraindicated. Do
not touch defective, open devices. Liability cannot be
accepted for damage resulting from misuse or failure
to observe the Operating Instructions.
– The user is responsible for testing Bluephase Style 20i
for its use and suitability for the intended purposes.
This is particularly important if other equipment is
used in the immediate vicinity of the curing light at
the same time.
– Use only original spare parts and accessories from
Ivoclar Vivadent (see Accessories). The manufacturer
does not accept any liability for damage resulting
from the use of other spare parts or accessories.
– The light probe is an applied part and may warm up to
a maximum of 45 °C (113 °F) at the interface to the
handpiece during operation.
Operating voltage
Before switching on, make sure that
a) the voltage indicated on the rating plate complies
with the local power supply and
b) the unit has acquired the ambient temperature.
If the battery or power pack are used separately,
e.g. during start-up or Click & Cure corded operation,
contact with patients or third parties must be
prevented. Do not touch the exposed contacts of
the connection plug (power pack).
Rating plate on the charging base
Assumption of impaired safety
If it has to be assumed that safe operation is no longer
possible, the power must be disconnected and the
battery removed to avoid accidental operation. This
may be the case, for example, if the device is visibly
damaged or no longer works correctly. A complete
disconnection from the power supply is only ensured
when the power cord is disconnected from the power
source.
Eye protection
Direct or indirect exposure of the eyes must be
prevented. Prolonged exposure to the light is
unpleasant to the eyes and may result in injury.
Therefore, using the supplied anti-glare cones is
recommended. Individuals who are sensitive to light,
who take photosensitizing drugs, have undergone eye
surgery, or people who work with the apparatus or in its
vicinity for long periods of time should not be exposed
to its light and wear protective orange goggles that
absorb light below a wavelength of 515 nm. The same is
true for patients.