Starting the Calibration Procedure
Lower the Empty Forks to the Ground
There should be no hydraulic pressure in hydraulic lift circuit.
- Turn power switch to on position
- If the LED display does not show any number(s), turn
on
the main power switch
located on top of the housing.
- LED display will show software version on the right side and number
8
will be
shown in
Mode
window.
Calibration with Empty Forks
With lowered empty forks on the ground and
Mode 8 showing
To initiate calibration procedure, press the
‘M’
button using a paper clip
Hold the ‘M’ button down for approximate 5 seconds.
After 5 seconds the
Mode 8
digit will change to
Mode 0
.
Press and hold the lift button to initiate the automatic weighing
cycle. Continue to hold until the system lift interrupt stops the lift
range. The system will develop an algorithm calculation and
determine the correct height to stop the lift cylinder.
The LED display will go blank momentarily, internal audio will be activated during
the measurement event and within few seconds the system will show a ‘
0’
value
in furthest right digit display. Audio indicator will be deactivated.
The ED2E-EP / DL2-EP zero weight of the empty fork calibration function is
complete