WatchPAT
™
ONE System
50
Operation Manual
APPENDIX D: SPO
2
ACCURACY IN THE WATCHPAT
The WatchPAT device uses Itamar Medical Pulse Oximetry system for the measurement of
functional oxygen saturation of arterial haemoglobin (SpO
2
). This appendix includes
information regarding the accuracy of these measurements following a clinical study of Itamar
Medical Pulse Oximetry.
1.
Overall, the Arms is estimated to be 1.9 for the range 70-100%
2.
The next table shows SpO
2
Accuracy Results:
Comparison to Reference CO-Oximetry
WatchPAT
* 70—100
90--100
80--<90
67--<80
A
RMS
Spec 3% for
range of 70-100%
# pts
1350
415
460
475
Pass
Bias
0.4
-0.4
0.6
0.9
A
RMS
1.88
1.10
1.62
2.54
* Note: The range of 70% to 100% includes reference data down to 67%
3.
The next plot shows the Bland-Altman plot for Itamar-Medical WatchPAT :
-20
-15
-10
-5
0
5
10
15
20
65
70
75
80
85
90
95 100
W
P
30
0
-
Re
f CO
-Ox
im
et
er
(Sp
O₂
-
SaO
₂)
Reference CO-Oximeter & WP300 (SaO2+ SpO₂ )/2
Bland-Altman Plot: WP300
Non-motion, 11 Subjects, December 13-15, 2017