Table 2 - from IEC 60601-1-2:2014
Guidance and manufacturer's declaration
– electromagnetic immunity – WatchPAT
The
WatchPAT
is intended for use in the electromagnetic environment specified below; The customer or the user of the
WatchPAT
should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2
Test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
8 kV contact
2 kV, +4 kV, + 8 kV,
+15 kV air
Not Applicable
2 kV, +4 kV, + 8 kV,
+15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%. Applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical public
low-voltage power supply network that
supplies buildings used for domestic
purposes, commercial or hospital, clinic
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Table 4 - from IEC 60601-1-2:2014
Guidance and manufacturer's declaration
– electromagnetic immunity – WatchPAT
The
WatchPAT
is intended for use in the electromagnetic environment specified below; The customer or the user of the
WatchPAT
should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2
Test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3V
0.15-80 MHz
Outside ISM Bands
6V
0.15-80 MHz
Inside ISM Bands
Not applicable
Not applicable
Portable
and
mobile
RF
communications
equipment
should be used no closer to any part of the
WatchPAT
, including cables, then the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter
Recommended separation distance