9
•
C
–
The indicator light provides the status of the Active Probe as follows:
o
Permanently green - Powered On and ready to use
o
Blinking green
–
Establishing connection with the BlueBox
o
Blinking blue
–
Reprogramming SPLASH
o
Permanently magenta
–
Golden image loaded and ready to use
•
D
–
FNet connector, that connects the Active Probe to the iC5700 BlueBox. The FNet cable is
delivered with the Active Probe.
When powering on the system, switch the iC5700 on prior to powering on the target
system; when powering down the system, power off the target microcontroller before
powering off the iC5700.
Use only original iSYSTEM accessories for powering and connecting with the iC5700.
Consult with iSYSTEM before attempting to use any other accessory.
Target connector
The target should feature a matching part, for example, Samtec part number: ASP-137969-01
(Samtec Series ERF8, Rugged High-Speed Socket).
Infineon TriCore target can also feature a standard 10-pin 1.27mm DAP target debug connector
exposing only the debug interface without the AGBT interface. The Active Probe can be used to
connect to such targets through a small adapter connecting at the end of Samtec 22-pin ERF8
connector.
10-pin 1.27 mm DAP Adapter
The 10-pin connector must be ordered separately in order to connect to the target featuring 10-pin
1.27mm pitch DAP target debug connector.
The following pinout is valid on the target side:
Signal direction
Signal
Pin Pin
Signal
Signal direction
I
VREF
1
2
DAP1
O / IO
Ground
GND
3
4
DAP0
O / O
Ground
GND
5
6
DAP2
I
Not Connected
7
8
DAP_EN
O
Ground
GND
9
10
~PORST
IO
10-pin 1.27mm TriCore DAP pinout
Ordering code IASAM22TRICOREP10-1