<Flow Set Command>
SR1 ~ SR8
- Set Range
Return Value : OK or ?!
Parameter : 1 <= MFC <= 5000 or 1 <= CM <= 50000
Parameter means RANGE Value.
If MFC Card is installed, MAX Range is 5000 but CM is 50000
※ <Caution> ※
If Range is changed, Set-point is changed to Zero.
EX)
<Host> 'SR1 1000<CR>'
<GMC1100> 'OK<CR>'
CH1 Range set by 1000
SS1 ~ SS8
- Set Set-point
Return Value : OK or ?!
Parameter : <= Range
Parameter means Set-point Value. Value is Range or less
EX)
<Host> 'SS1 1000<CR>'
<GMC1100> 'OK<CR>'
CH1 Set-point set by 1000
SF1 ~ SF8
- Set Flow ON/OFF
Return Value : OK or ?!
Parameter : 0 ~ 1
Parameter means that 0: Off, 1:On
EX)
<Host> 'SF1 1<CR>'
<GMC1100> 'OK<CR>'
It is CH1 Flow On.