STERILIZZATORE D’ARIA
HUNTER UVC
Other technical-construction characteristics of the product
Characteristic
References
Code for spare parts
6 x 55W UVC lamps for each module
Type 55W (double-ended)
MPI17623
UVC corner
9000 hours (about 1 year in continuous
service)
N ° 3 electronic ballasts with PFC - 2x58W
Type VS ELXc 258.210 n° rif188707
MPI17642
UVC wavelength
253,7 nm
Ultraviolet energy x each lamp
2,1 W/m² at the distance of 1 mt
Electronic timer K400-2M
K400-2M 200-240Vac
MPI17618
Internal hour meter that cannot be reset
MPI17603
Operating temperature
-10 °C + 55°C
ELECTRICAL CHARACTERISTICS
Model
Single-phase supply
voltage
[Vac]
Electricity frequency
[Hz]
Electric power max.
[W]
HUNTER UVC
230
50
350W
ELECTROMAGNETIC FIELDS
Model
Maximum measured weighted
value
Maximum limit imposed
(CEI EN 62233)
outcome
Hunter UVC
0,07
1
COMPLIANCE
Storage conditions
Temperature -5 ° C / 45 ° C
Humidity 10 - 75%
Pressure 500-1060 hPa
Before proceeding with the installation, check the integrity of the device; if there are signs of
damage, do not operate the appliance.