S p e z i a l e l e k t r o n i k G m b H
iseg Spezialelektronik GmbH
Email: [email protected]
Tel ++ 49 (0) 351 / 26 996 - 0
Bautzner Landstr. 23
http://www.iseg-hv.de
Fax ++ 49 (0) 351 / 26 996 - 21
D - 01454 Radeberg/ Rossendorf
9
2.2
Technical data device class HPS, 800 W
Table 2.3: Technical data HPS 800W
Device class HPS, 800 W
Output power P
nom
[W]
800
Polarity
Factory fixed, n negative or p positive
Efficiency
> 85% (V
in
= 230 V)
Option capacitor charger (CLD)
Very low output voltage overshoot
Ripple and noise
Standard
∆
v
out
< 0.01%
∗
V
nom
(> 10 Hz), V
out
≤
8 kV
∆
v
out
< 0.05%
∗
V
nom
(> 10 Hz) V
out >
8 kV
Option CLD
∆
v
out
< 1.5%
∗
V
nom
(> 10 Hz)
Stability
∆
v
out
< 0.05%
∗
V
nom
(for 8 h with constant conditions, after ½ h warmup)
Voltage regulation
∆
v
out
< 0.01%
∗
V
nom
(
∆
v
in
, 0
≤
I
out
≤
I
nom
, 5 V
≤
V
out
≤
V
nom
)
Current regulation
∆
I
out
< 0.2%
∗
I
nom
(
∆
v
in
, 5 V
≤
V
out
≤
V
nom
)
Accuracy
Voltage:
< 0.5% • V
OUT
+ 0.3%
·
V
NOM
for one year
Current:
< 0.1% • I
OUT
+ 0.05%
·
I
NOM
for one year
Temperature coefficient
< 2
∗
10
-4
/K
Front panel control
Front panel operation via rotary encoders and displays (LCD)
Remote control
Option AIO
Analogue signals
Level
0 V – 5 V
Digital signals
Low level
0 V – 1 V
High level
3.5 V – 5 V or open
Digital interface
USB, CAN, Option: RS232 (RS2), Ethernet (ETH), IEEE 488 (IEE)
Supply
V
in
= 85 V – 264 V AC (PFC)
I
in
= 4.5 A (V
in
= 230 V, P
nom
); I
in
= 9 A (V
in
= 115 V, P
nom
)
Line frequency 47 Hz < f
l
< 63 Hz
Internally fused with 2 x 10 A with slow characteristic
Inrush current approx. 25 A
External fuse
10 A, with slow characteristic
Cooling
Forced cooling with integrated fans (
≤
20 m³/h)
Monitoring
Single phase mains voltage, auxiliary voltage, over voltage, temperature
Working conditions
Temperature:
0°C to 35 °C
Humidity:
20% to 90%, no condensation
Storage conditions
Temperature:
-25°C to 80 °C
Humidity:
20% to 90%, no condensation
Electromagnetic
compatibility
Emission
EN 55011 (curve B)
Immunity
EN 61000 4-2, EN 61000 4-3, EN 61000 4-4, EN 61000 4-8
Safety standard
EN 61010-1 (VDE 0411)
Maximum complete discharges (ARC) per time
ARC Management
See 3.3 ARC Management
Voltage ramp, (fastest volt-
age ramp, slower ramp
speeds can be programmed)
Standard
3000 V/s
Option CLD
> 10
∗
V
nom
/ s
Electrically isolated return of the high voltage
Potential difference between return conductor and protective ground up to ± 400 V