2.1 General technical data
19“ / 1U - series
HPx
)1
300 W
19“ / 1U - series
HPx
)1
800 W
19“ / 1U – series
LPx
)1
300 W
19“ / 1U – series
LPx
)1
800 W
Accuracy
voltage measurement
± (0,5%
•
V
OUT
+ 0,3%
•
V
NOM
+ 1 digit)
for one year
current measurement
± (0,1%
•
I
OUT
+ 0,05%
•
I
NOM
+ 1 digit)
for one year
Temperature coefficient
< 2
•
10
-4
/
K
Display
8-digit LCD-Display for current and voltage
Resolution of voltage and current
measurement
4-digit
Resolution of settings
LOCAL 4-digit
Voltage / Current
REMote 4-digit
Switching of output voltage
with button “ON/OFF” or via remote control
Control
LOCAL
rotary encoders for voltage and current
(REMote)
CAN
via CAN-Interface
USB
via USB-Interface
optional:
RS-232
via RS232-Interface
optional:
AIF
via indirect coupled analogue I/O additionally (male SUB-D-9)
optional:
IEEE-488
via IEEE-Interface additionally (male Micro D25)
optional:
ETH
via Ethernet-Interface additionally (RJ45)
Efficiency
up to 85%
Supply
V
IN
= 85 up to 260 V AC with PFC
I
IN
= 1,7 A at 230V-AC / 3,5 A at 115 V-AC (300 W series),
I
IN
= 4,5 A at 230 V AC / 9 A at 115 V AC (600 W series)
via mains connector and switch “POWER“, isolated from HV-output,
fused with 2
•
6,3 A / slow (300 W series)
2
•
10 A / slow (800 W series).
Dimension
1U -19“ compatible / depth: 410 mm
Weight
ca. 5,7 kg (300 W series) / ca. 6,5 kg (800 W series)
Cooling
Internal fan
Protection
Over load and short circuit , voltage supply and temperature
Environment conditions
Operating temperature:
5 up to 35 °C
Humidity:
30% up to 80 %, no condensation
Storage temperature
0 up to 60 °C