Channel event status (read/write access)
:READ:CHANnel:EVent:STATus?
Bit15
Bit14
Bit13
Bit12
Bit11
Bit10
Bit9
Bit8
EVLIM
ECLIM
ETRP
EEINH
EVBNDs
ECBNDs
res
res
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2
Bit1
Bit0
ECV
ECC
EEMCY
EEOR
EOn2Off
EIER
res
Res
The Channel EventStatus register describes the
captured
status. Depending on the status of the
channel the bits will be set but not reset. A reset has to be done by the user by writing an 1 to this
event bit. All Events can be cleared by :EVENT:CLEAR.
Bit
Name
Description
EVLIM
EventVoltageLimit
Event: Voltage limit has been exceeded
ECLIM
EventCurrentLimit
Event: Current limit has been exceeded
ETRIP
EventTrip
Event: Trip is set when Voltage or Current limit or Iset has been ex-
ceeded (when KillEnable=1)
EEINH
EventExtInhibit
Event: External Inhibit
EVBNDs
EventVoltageBounds
Event: Voltage out of bounds
ECBNDs
EventCurrentBounds
Event: Current out of bounds
ECV
EventControlledVoltage
Event: Voltage control
ECC
EventControlledCurrent
Event: Current control
EEMCY
EventEmergencyOff
Event: Emergency off
EEOR
EventEndOfRamp
Event: End of ramp
EOn2Off
EventOnToOff
Event: Change from state "On" to "Off" without ramp
EIER
EventInputError
Event: Input Error
res
Reserved
An event bit is permanently set if the corresponding status bit is 1 or is changing to 1. Different to the
status bit an event bit isn't automatically reset. A reset has to be done by the user by writing an 1 to
this event bit.
If one of the EventStatus Bits EVLIM, ECLIM, ETRIP, EEINH, EVBNDs, ECBNDs, EEMCY is set, it
prevents turning on the High Voltage again until the corresponding bit is cleared.