4.2 Menu
In HV-OFF mode the device menu is activated by pressing the button MENU.
If no button is pressed, the display switch back to HV-OFF mode after 30 seconds. The menu can
also be closed without changing any value by pressing the button ESC.
By turning the rotary encoder VOLTAGE you can scroll through the menu. By pressing the rotary en-
coder VOLTAGE the displayed menu point is selected. The setting can be changed by the active ro-
tary encoder (shown by yellow LED). By pressing the active rotary encoder the changes are stored
and the main menu is displayed again.
Display
Description
F01 Set Limit V
Adjust the Software voltage limit V
OUTmax
in the range of 0.02
•
V
NOM
≤
V
OUTmax
≤
V
NOM
with
rotary encoder VOLTAGE. V
SET
will be limited to this value.
If the Limit gets smaller than the actual value of V
SET
, V
SET
will be decreased to V
OUTmax
automatically.
F02 Set Limit I
Adjust the Software current limit I
OUTmax
in the range of 0.02
•
I
NOM
≤
I
OUTmax
≤
I
NOM
with
rotary encoder CURRENT. I
SET
will be limited to this value.
If the Limit gets smaller than the actual value of I
SET
, I
SET
will be decreased to I
OUTmax
auto-
matically.
F03 Set Ramp V
Adjust voltage ramp speed with rotary encoder VOLTAGE in the range of 1…3000 V/s
(factory setting is 0.2 • V
NOM
/s).
On LPS devices, the software voltage ramp can be disabled. In this case, the device
changes it’s output voltage as fast as possible.
The software ramp can be disabled be setting the ramp speed greater than 3000 V/s:
“max.” will be displayed. To enable the software ramp again, set a ramp speed between
1…3000 V/s.
F04 Set Ramp I
Adjust current ramp speed with rotary encoder CURRENT with the given values in the
range of 0.02 • I
NOM
/s up to I
NOM
/s (factory setting is I
NOM
/s)
F05 Auto Start
Not implemented yet.
F06 Auto AIF
AIF ON:
Turn on HV by pushing the ON/OFF button or with
INHIBIT Low to High on analogue I/O
AIF OFF
Turn on HV by pushing the ON/OFF button only
The INHIBIT signal on analogue I/O has priority in both cases!
INHIBIT
High to Low:
turn off HV
Low to High:
turn on HV (KILL disable)
Low static:
HV = 0
F07 Set Interfce
Select external Interface with rotary encoder VOLTAGE:
“CAN”
control via CAN-Interface
“RS-232”
control via RS232-Interface
“USB“
control via USB-Interface
“IEEE 488” control via IEEE (GPIB)-Interface
“Ethernet“ control via Ethernet-Interface
“AIF”
control via Analogue I/O
F08 Set Instruct
Select instruction type for RS-232/USB/IEEE-488/Ethernet control with rotary encoder
VOLTAGE:
”EDCP”
SCPI command set with EDCP (recommended)
“SCPI”
old SCPI command set
“ET”
old ET command set
F09 Addr IEEE
Select IEEE address with rotary encoder VOLTAGE: 01 to 30. Factory setting is 17.
F10 Addr CAN
Select CAN address with rotary encoder VOLTAGE: 00 to 63. Factory setting is 0.