Read out the measurement values
Read measured output voltage:
STATUS,MU
UM,
RANGE=3.000kV,
VALUE=2.458kV
Read measured output current:
STATUS,MI
IM,
RANGE=100.0mA,
VALUE=25.3mA
Read out the device status
STATUS,DI
DI,
b
15
b
14
b
13
b
12
b
11
b
10
b
9
b
8
b
7
b
6
b
5
b
4
b
3
b
2
b
1
b
0
0
1
#define IERR
b15
no Input Error
Input error
#define RAMP
b14
no Ramp
Ramp
#define EMCY
b13
no Emergency
Emergency
#define TRIP
b12
no Trip error
Trip error
#define RES5
b11
Reserved
#define RES4
b10
Reserved
#define RES3
b9
Reserved
#define RES2
b8
Reserved
#define SERR
b7
no Sum error
Sum error
#define IREG
b6
no Current control
Current control
#define VREG
b5
no Voltage control
Voltage control
#define POL
b4
negative Polarity
positive Polarity
#define INH
b3
no external Inhibit
external Inhibit
#define RES1
b2
Reserved
#define KILena
b1
Kill disable
Kill enable
#define ON
b0
no High voltage active
High voltage active
Read out the LAM status
STATUS,LAM
LAM,ERROR
External Inhibit occurred during Kill enable
LAM,EMERGENCY
High Voltage was shut down with Emergency Off
LAM,SAFETY LOOP
Safety Loop was opened
LAM,INHIBIT
External Inhibit occurred
LAM,TRIP ERROR
Software current trip occurred
LAM,VOLTAGE LIMIT
Voltage has exceeded set value
LAM,CURRENT LIMIT
Current has exceeded set value
LAM,INPUT ERROR
Wrong command received
LAM,OK
Status OK
Read out the device identification
ID
ID,
iseg
Spezialelektronik
r5.01
sn.680041
Typ HPn
30