GLS Sampler
3-14
inside the liquid detector and pump, and that the tube is
not kinked where it fits over the bulkhead fitting.
Note
The factory set value of 500,000 pump counts will
deliver approximately 500 samples of 200 ml each,
using a
3
/
8
-inch by 10-foot suction line at a 5-foot suction
head.
Note
The peristaltic pump and tube will perform the best
when you:
- Use Isco replacement pump tubes or bulk tubing.
- Install the tube properly, aligning the inside edges of
the bands with the outside edges of the liquid
detector.
- Follow the natural curve of the pump tube when fitting
the tube inside the pump housing.
- Use the shortest possible length of suction line.