Section 2 Programming
2-13
2.4.7 Suction Line
The display alternates between:
This Suction Line display reports the suction
line diameter and length. This information
must be correct so that the GLS can deliver
sample volumes as programmed. Verify that
the diameter and length match the suction line
in use.
If the suction line settings are correct,
calibration is not necessary. To skip the
calibration, press the Go button to run the
program, or press the Enter button to accept
the values and return to the Standby state.
If the settings do not match the suction line in
use, you must enter new values. Press the
Calibrate button and proceed with the
Calibration steps (see section 2.7).
While the two displays alternate, the GLS is
counting down a five-minute time interval. If
you do not press a button during this count
down, the GLS times out and automatically
runs the current program.
If your sampling protocol requires the utmost
volume accuracy, step through the calibration
SUCTION LINE
9mm BY 7.6 m
PRESS GO TO RUN OR
CALIBRATE TO CHANGE
and