GLS Sampler
2-18
6. The GLS goes through a complete sample
collection cycle and deposits the sample into
the graduated cylinder.
7. When it has finished collecting the sample,
the GLS displays the expected volume.
Measure the volume in the graduated
cylinder. If the actual and displayed
amounts differ, enter the actual amount
using the number-entry buttons. Press
Enter to accept the value.
8. If the actual amount that you enter is
significantly different from the expected
volume, the GLS will ask you confirm your
entry. Re-measure the liquid in the cylinder
and compare it with the displayed value. If
they match, select YES and the GLS
updates its internal pump tables and
returns to the Standby state. If they do not
match, select NO and repeat step 5.
Note
Repeat the calibration steps to check volume accuracy.
TAKING 200 ml
CALIBRATE SAMPLE
VOLUME DELIVERED
200 ml
___ml! ARE YOU
SURE? YES NO
Содержание GLS