background image

18

OcuLight Symphony Operator Manual

31144C-EN

UL

LISTED

86Z4

MEDICAL

EQUIPMENT

C

11

5

Foot

Switch

STOP

Remote

Interlock

4

P

U

S

H

R

E -CI

R

K

-

I

T

LE LASER RADIA

TION

N EXPOSURE T

O

ERED RADIA

TION

DUCT (EN60825-1:1998)

ODUCT 21 CFR 1040

Po = 2 mW

= 630-670 nm

If desired, connect a remote interlock switch

As a safety feature, a remote interlock switch may be installed, for example
a switch that disables the laser whenever a treatment room door opens
while the console is in the Treat mode. If you will be using a remote
interlock switch, install a remote interlock switch to each console as
described:

Connecting to the OcuLight SLx console

Attach the two wires connecting the door switch to the remote interlock
plug in place of the shorting connection.

Insert the remote interlock plug (either by itself or attached to a remote
interlock switch) into the remote interlock port.

If the remote interlock plug is not properly connected, when you turn on
the OcuLight SLx console, 

No Remote Interlock

 displays on the console

status panel.

Connecting to the OcuLight GLx/TX console

Remove the factory installed wire and in its place attach the two wires
leading to your own switch.

If the remote interlock switch is not properly connected, when you turn on
the OcuLight GLx or TX console, 

No Remote Interlock

 displays on the

console status panel.

Содержание OcuLight Symphony

Страница 1: ...OcuLight Symphony IRIDEX 31144C EN Operator Manual...

Страница 2: ...s as were published with the original manual This manual is provided for instructional use and while every effort has been made to ensure that the information is accurate the text illustrations and sp...

Страница 3: ...console 2 Treatment wavelength 2 Spot size 2 IRIDEX SLIT LAMP WORKSTATION SPECIFICATIONS 3 Microscope 3 Type 3 Magnification 3 Oculars 3 Diopter adjustment 3 Field of view 3 Interpupillary adjustment...

Страница 4: ...ation only 21 Installing the OcuLight Symphony Adapter 25 USING THE SLIT LAMP AND TABLE CONTROLS FOR IRIDEX SLIT LAMP WORKSTATION ONLY 34 Turning the slit lamp illumination fixation lamp on off 34 Adj...

Страница 5: ...LY 54 REPLACING THE SLIT LAMP FIXATION LIGHT BULB FOR IRIDEX SLIT LAMP WORKSTA TION ONLY 55 CHANGING THE SLIT LAMP FUSE FOR IRIDEXSLIT LAMP WORKSTATION ONLY 56 CLINICAL SAFETY 57 CLINICAL APPLICATIONS...

Страница 6: ...iv OcuLight Symphony Operator Manual 31144C EN...

Страница 7: ...or precise focus and consistent burns superior viewing through a dual wavelengh eye safety filter featuring UltraView optics which protects from both infrared and green wavelengths and a switch that e...

Страница 8: ...on Zeiss 30 SL Zeiss 30 SL M Standard spot 810 nm for the OcuLight SLx console Treatment wavelength Laser diode 810 nm infrared Spot size 125 200 350 600 1000 m all parfocal Large spot 810 nm LS for t...

Страница 9: ...ment 6 diopter Field of view 5 2 8 14 24 37 mm Interpupillary adjustment 48 5 to 80 mm Stereoscopic angle 8 Slit illumination Slit width 0 to 14 mm continuously variable Slit length 14 mm 1 8 12 conti...

Страница 10: ...e 0 horizontal Free working distance 6 9 cm from exit prism to patient s eye Joystick movement Z vertical 35 mm X axis 10 8 cm Y axis 11 3 cm Horizontal fine movement 10 mm Slit lamp table top dimensi...

Страница 11: ...registration Please complete and forward to us the enclosed product registration cards Service and technical support IRIDEX has established an efficient process to support its installations worldwide...

Страница 12: ...6 OcuLight Symphony Operator Manual 31144C EN...

Страница 13: ...er photocoagulation NOTE Should you notice problems with your order please contact your local IRIDEX Technical Support representative immediately REFERENCE See your OcuLight console manuals for descri...

Страница 14: ...s LIO and the Endoprobe handpieces It includes various cable guides to prevent damage to the fiber optic assembly footswitch and power cables Slit lamp table for the IRIDEX slit lamp workstation only...

Страница 15: ...OcuLight Symphony Adapter It consists of a patient chinrest and headrest with fixation light a viewing assembly with oculars magnification changer and objective lens an illumination prism an illumina...

Страница 16: ...e OcuLight Symphony Adapter attaches to the slit lamp enabling the slit lamp to be used as the therapeutic device The adapter has an integral fiber optic assembly which houses up to three fiber optic...

Страница 17: ...pulator adds precision beam steering capabilities to the OcuLight Symphony Adapter while offering a full range of motion over the treatment area The finger rest provides a stable comfortable resting p...

Страница 18: ...anel from the cart REFERENCE See your OcuLight console manuals for important information about choosing the location and position for the consoles Information IRIS Medical OcuLight SLx POWER DURATION...

Страница 19: ...ng the console components Connect the main power cables Thread the main power cables through the hole on the back of the shelf and again through the hole on the bottom shelf of the components cart Com...

Страница 20: ...lug the other power cable into the OcuLight GLx console AC power inlet OcuLight GLx TX console power cable OcuLight GLx TX console OcuLight SLx console OcuLight SLx console power OcuLight GLx console...

Страница 21: ...hole in the bottom bin on the components cart and again through the hole on the bottom shelf of the components cart NOTE The OcuLight TX console will require the use of a 2nd footswitch The footswitch...

Страница 22: ...enter the Treat mode Connect Footswitch displays on the console status panel Connect the GLx footswitch jumper cable OcuLight GLx console OcuLight TX console NOTE The OcuLight TX console will require...

Страница 23: ...y connected when you turn on the console and enter the Treat mode Connect Footswitch displays on the console status panel Footswitch jumper cable five pin end Footswitch jumper cable four pin end NOTE...

Страница 24: ...o each console as described Connecting to the OcuLight SLx console Attach the two wires connecting the door switch to the remote interlock plug in place of the shorting connection Insert the remote in...

Страница 25: ...put contact If desired connect an external warning device As an additional safety measure you may connect an external warning device such as a warning light or small rotating beacon to the Auxiliary O...

Страница 26: ...Clip the footswitch and power cables into the cable guides Cable guide Channel Attach the back panel to the cart Install the cable guides for the fiber optic assembly To keep the fiber optic cables o...

Страница 27: ...screws 4 If a small metal cover is attached to the underside of the base remove and reserve the screws from each end The underside of the cover is attached to a grounding wire by a star washer and fi...

Страница 28: ...he base raise the base and place it on top of the pedestal 8 Balancing the opposite end of the base on a chair or table route the cables from the pedestal through the hole in the base 9 Secure the ped...

Страница 29: ...to the holes on the arc side of the base Screw the other two wheels to the opposite corners of the base 13 Turn the pedestal and base upright 14 Plug one end of the power cable into the base inlet plu...

Страница 30: ...of the table top secure the table top to the pedestal with the four screws 0 0 I 17 Connect the small electrical cable from the pedestal to the slit lamp power supply 18 To keep the fiber optic cables...

Страница 31: ...0 1 2 8 20 5 1 2 32 Install the IRIDEX split mirror illumination prism for Zeiss 30 SL and 30 SL M slit lamps only Replace the original illumination prism with the IRIDEX split mirror illumination pr...

Страница 32: ...the lower edge into the dovetail mount 6 Tilt the eye safety filter upright and then place the top edge into the dovetail mount Use the set screw provided with the OcuLight Symphony Adapter and tight...

Страница 33: ...e eye safety filter and insert the lower edge into the dovetail mount 5 Tilt the eye safety filter upright and then place the top edge into the dovetail mount Use the set screw provided with the OcuLi...

Страница 34: ...back into its centered position 0 15 30 45 60 75 90 105 120 1 2 8 5 1 2 32 Attach the micromanipulator handle Insert the micromanipulator handle into the OcuLight Symphony Adapter socket and tighten t...

Страница 35: ...optic cable could cause accidental laser exposure or injury to yourself your patient or others in the treatment room NOTE Avoid making any sharp bends or kinks in the cable assembly as this can compro...

Страница 36: ...lamp table Secure the fiber optic assembly to the components cart Using the cable guides installed previously on the components cart secure the fiber optic assembly to the slit lamp table Fiber optic...

Страница 37: ...Light SLx console fiber optic port until secure If the fiber optic is not properly connected when you turn on the OcuLight SLx console Connect Fiber displays on the console control panel TERVAL COUNTE...

Страница 38: ...ve caps from the OcuLight GLx TX console fiber optic port and from the fiber optic connector Carefully insert and finger tighten the connector into the OcuLight GLx TX console fiber optic port until s...

Страница 39: ...e SmartKey port For example if you will be using the OcuLight SLx console for treatment insert the SmartKey into the OcuLight SLx console SmartKey port If present store the third unused fiber optic If...

Страница 40: ...to occur to reduce discomfort for the patient always keep the illumination intensity at the minimum required for treatment Adjusting the illumination intensity Use the slit lamp illumination control...

Страница 41: ...towards the patient Turn the control clockwise to raise the slit lamp and counter clockwise to lower When the slit lamp is in the desired position turn the base tension knob clockwise to secure the po...

Страница 42: ...nterpupillary distance for your eyes 1 2 8 2 0 5 1 2 32 X Serial Made in USA SCxxxx 8 2 0 5 1 2 X Serial Made in USA SCxxxx Focus the oculars Rotate the oculars to your approximate prescription Set th...

Страница 43: ...he oculars and magnification housing in place 0 15 30 45 60 75 90 105 120 1 2 8 20 5 1 2 32 Tension control knob If desired set the slit lamp for off axis illumination Loosen the tension control to un...

Страница 44: ...75 90 105 120 1 2 8 20 5 1 2 32 Slit height control dial NOTE To avoid discomfort and protect the patient s eyes from excessive optical radiation reduce the size of the illuminated area slit height an...

Страница 45: ...the dial by a colored marker cobalt blue filter blue marker red free filter green marker heat absorbing red marker white marker no filter 0 15 30 45 60 75 90 105 120 1 2 8 5 2 Filter control dial NOT...

Страница 46: ...nm 810 LS large spot 810 nm and 810 nm standard spot 810 nm REFERENCE See the Operations section of your OcuLight console manuals for instructions on using the console controls REFERENCE Read the Clin...

Страница 47: ...post if using the IRIDEX slit lamp or position a paper target on the headrest if using a Zeiss 30SL or 30 SL M slit lamp 2 Set the oculars to the physician s approximate prescription 3 Turn on the sli...

Страница 48: ...ng the spot 9 Turn on the OcuLight console and place it in Treat mode 10 Use the lever labeled X to adjust the horizontal position of the laser spot into the center of the slit Serial Made in USA SCxx...

Страница 49: ...eye safety filter protects you from laser radiation above regulatory requirements Opening the filter while using the OcuLight GLx TX console When the filter is open while using the OcuLight GLx consol...

Страница 50: ...fiber optic connect the fiber optic labeled 810 LS and the SmartKey to the OcuLight SLx console If using the OcuLight SLx console for treatments requiring the standard spot fiber optic connect the fi...

Страница 51: ...ptic with the Fiber Switch on the Symphony Slit Lamp Adapter Resume treatment following steps 7 11 above 13 When you do not require the treatment beam remove your foot from the footswitch and place th...

Страница 52: ...thorized use 4 Handle the contact lens according to the manufacturer s instructions 5 If desired clean the OcuLight Symphony Adapter fiber optic connectors and replace the protective caps If desired c...

Страница 53: ...tive Problem Action Inadequate or no aiming beam Ensure that the OcuLight Symphony Adapter is properly connected to the console Ensure that the OcuLight SLx GLx TX console is in Treat mode Ensure that...

Страница 54: ...DEX Technical Support representative The Status panel reads Connect Fiber Ensure that the OcuLight Symphony Adapter is properly connected Connect SmartKey Ensure that the SmartKey is properly inserted...

Страница 55: ...n The Status panel reads Unknown Fiber Type Connect the 532 nm fiber optic connector to the OcuLight GLx or TX console Incompatible Device Use with 810 nm console Connect the 810 nm or 810 LS fiber op...

Страница 56: ...50 OcuLight Symphony Operator Manual 31144C EN...

Страница 57: ...r bend the fiber optic cable When connected to the console ensure that the fiber optic cable is located away from high traffic areas Keep the delivery mirror free of fingerprints Inspect the OcuLight...

Страница 58: ...ilter 1 Wrap a lens tissue around one end of a cotton tipped swab 2 Place several drops of 100 ethanol 100 methanol or high grade acetone on the tissue 3 Wipe the lens gently with the swab to remove a...

Страница 59: ...any cleaning agents on the slit lamp Clean the optical components Use a soft brush to remove dust from accessible surfaces When necessary clean the front surface of the objective carefully 1 Turn on...

Страница 60: ...d pulling down the door 3 Ensure that the bulb has cooled 4 Remove the old bulb 5 Hold and properly orient a new bulb by the plastic envelope not by the bulb insert the new bulb into the bulb housing...

Страница 61: ...slit lamp main power cable from the electrical outlet 2 Remove the red cap from the fixation point 3 Remove the defective light bulb 4 Insert a new light bulb 5 Replace the red cap on the fixation li...

Страница 62: ...is the fuse cartridge 3 Using a small screwdriver gently push against the two locking tabs on either side of the fuse cartridge until the cartridge pops up from the port 4 Remove the cartridge 5 Remo...

Страница 63: ...is not intended to suggest treatment techniques Qualified physicians should review the available literature presented in clinical papers before using the OcuLight Symphony Adapter Power density and s...

Страница 64: ...ny other console Preventing reflection hazards Laser beams reflected from specular surfaces can harm your eyes the patient s eyes or others eyes Any mirror or metal object which reflects the laser bea...

Страница 65: ...Center for Devices and Radiological Health of the Food and Drug Administration FDA CE labeled devices comply with all appropriate performance standards as specified in Annex II of the Medical Device...

Страница 66: ...EN Serial number label LASER APERTURE Y CE label Fiber Switch label Laser aperture laser emission labels Eye safety filter label Serial Made in USA SCxxxx LASER APERTURE 532 nm 810 LS Location of reg...

Страница 67: ...1 Clinical Safety 31144C EN IRIS Medical Ser Nr XXXXXXX Model SL980 5 Special 6V 3 5A 50 60HZ Distributed by IRIDEX Corporation Mfg C S O Italy On back of slit lamp base Slit lamp specifications label...

Страница 68: ...62 OcuLight Symphony Operator Manual 31144C EN...

Отзывы: