Troubleshooting
Reporting problems
Should you experience problems with your OcuLight Symphony system, refer
to the suggestions in this section. If you continue to experience problems, write
down the error message, product name, and serial number of the
OcuLight SLx/GLx/TX console, the OcuLight Symphony Adapter, and the
slit lamp before contacting your local IRIDEX Technical Support representative.
Problem
Action
Inadequate or no aiming beam
• Ensure that the OcuLight
Symphony Adapter is
properly connected to the
console.
• Ensure that the
OcuLight SLx/GLx/TX
console is in Treat mode.
• Ensure that the Fiber Switch
is in the correct position for
the console and wavelength
you are using.
• Turn the aiming beam
control fully clockwise.
• Ensure that the fiber-optic
connectors are not damaged.
• If the aiming beam is still not
visible, contact your local
IRIDEX Technical Support
representative.
REFERENCE
See the Troubleshooting
section of your OcuLight
console manuals for
additional information.