Problem
User Action(s)
The treatment lesions are variable
or intermittent (LIO only)
• The LIO may be slightly out of focus. This decreases power density.
Readjust your working distance to obtain the smallest spot size.
• A poorly centered laser beam may be clipping on the examination lens or
on the patient’s iris. Adjust the laser beam in the illumination field.
• The laser treatment parameters may be too close to the tissue response
threshold for consistent response. Increase the laser power and/or
exposure duration, or select a different lens.
Does not fit on the mounting plate
(OMA only)
• Inspect and clean the mounting plate.
• Verify that the mounting plate corresponds to your microscope.
Laser and viewing systems are
not focussed at the same point
(OMA* only)
• Verify installation of a 175 mm microscope objective lens on
the microscope.
• Turn on the aiming beam to determine focus position and adjust
as necessary.
View is blocked or partially
blocked by OMA (OMA* only)
Set magnification to 10X or more.
*Operating microscope adapter compatible with IRIDEX IQ 810 and SLx Systems.