Maintenance
These probes are single-use, disposable delivery devices. The device is intended for use on one
patient during a single procedure. It is not intended to be reprocessed (cleaned, disinfected/
sterilized) and used on another patient. No reuse processes have been validated by the manufacturer.
The reuse of a single-use medical device can compromise the safety and health of patients, users or
third parties. Reuse risks the introduction of contaminants and/or microbes and may lead to patient
injury and/or infection.
Store probe package under normal storage conditions - in a dry, clean, well ventilated area at room
temperature, between 15º-25ºC (59-77ºF).
Cleaning Optical Components
1. Wrap a lens tissue around one end of a cotton-tipped swab.
2. Place several drops of 100% ethanol, 100% methanol, or high-grade acetone on the tissue.
3. Wipe the lens gently with the swab to remove all dust and debris.
4. If the surface is still not clean, put a clean lens tissue around the end of the swab and gently wipe
it again.
Optical components refer to eye safety filters, output prism of DioPexy handpiece, mirrors on slit
lamp adapters, operating microscope adapters, etc.