Guidance and manufacturer’s declaration – electromagnetic immunity
NOTE: U
T
is the a.c. mains voltage prior to application of the test level
The ET-9090/ET-1313/ET-7070 is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the ET-9090/ET-1313/ET-7070
should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) mag-
netic field IEC
61000-4-8
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV line(s)
and neutral
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5s
Not applicable.
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV line(s)
and neutral
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5s
3 A/m
Electromagnetic environment
— guidance
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
ET-9090/ET-1313/ET-7070 requires
continued operation
during power mains interruptions, it
is recommended that the
ET-9090/ET-1313/ET-7070 be pow-
ered from an uninterruptible power
supply or a battery.
Not applicable.