 (Model # ET-9090, ET-1313, and ET-7070) Stimulator Technical Specifications

Channel: Dual, isolated between channels.

Pulse Amplitude: Adjustable 0 – 80mA peak into 500

Ω

 load each channel.

Pulse Rate: As pre-programming operation mode.

Pulse Width: As pre-programming operation mode.

Software Ramp Up Feature: Pulse width ramp up when changing mode.

Timer: 5~60 min. selectable.

LCD: Show modes, pulse rate, pulse width, timer, CH1/CH2, intensity level. 

Program Modes:

Wave Form: Symmetrical Bi-Phasic square pulse.

Max Charge per Pulse: 20.8 micro-coulombs maximum.

Power Source :  3 x AAA / 4.5 Volt batteries 

All electrical specifications are ±20% at 500

Ω

 load.

Description of Symbols: 

(i) There are a number of technical symbols on your iReliev

®

 unit explained as follows: 

Program

Max.

Phase duration

Rate

Function mode Wave form Type

P1

80mA

260uS

15Hz

Constant

A

P5

80mA

260~150uS

60Hz

Modulated

A

P2

80mA

260uS

60Hz

Modulated

A

P6

80mA

260uS

7 <->60Hz

Modulated

C

P3

80mA

260uS

60Hz

Constant

A

P7

80mA

260~156uS

60Hz

Modulated

A

P4

80mA

260~150uS

2 ~ 60Hz

Modulated

B

P8

80mA

P1 ~P7

Cycle

A/B/C

This symbols means “Serial number.“

This symbols means “Attention” consult the 
accompanying documents. 

This symbols means “Manufacturer.“ 

This symbol means “used before”, represented as “YYYY-MM” (for year and 
month). 

This symbol means type BF equipment; this device offers protection 
against electrical shock by standard compliance to leakage currents of 
electrode pad.

(ii) Package of electrode pads are labeled as 
follows:  

14

Содержание ET-9090

Страница 1: ...Instruction Operating Manual Read Before Using MODEL ET 9090...

Страница 2: ...le of Contents 2 Introduction 3 Safety Information 4 Battery Safety Information 5 Cautions Contraindications 5 Package Contents 6 Parts Diagram Features 7 Inserting Changing the Batteries 8 Putting on...

Страница 3: ...methods such as medication in jections surgical procedures or with alternative treatment methods such as acupuncture or TENS Electrotherapy treatment can block pain signals by interrupting the transm...

Страница 4: ...rs This may impair the device s ability to function Do not turn on the device until wrap has been snugly secured on the body Do not use the device while taking a bath or a shower Do not use this devic...

Страница 5: ...device Consult your doctor always before use if you suffer from epilepsy or certain disabilities People whose sensitivity is disturbed or impaired should only apply this device under supervision of y...

Страница 6: ...e iReliev system if you have undiag nosed chronic pain system if you have undiag nosed chronic pain DO NOT TOUCH ELECTRODE S AFTER TURNING ON When you are familiar with the operation of your iReliev d...

Страница 7: ...On Adjust Increase Key 6 Power Off Adjust Decrease Key 7 Program Mode Therapy Time Selection 8 CH 1 Channel 1 Key 9 CH 2 Channel 2 Key 10 Program Number 11 Therapy Time Remaining 12 CH 1 Intensity Lev...

Страница 8: ...artment make sure to match up the sym bols 3 Close the battery cover by carefully placing the stud into the slot in the rear area and sliding it upward applying slight pressure 4 Follow the same proce...

Страница 9: ...ow the waist The Wrap should fit very snugly around your body as follows 4 When the wrap has been placed in the correct position bring the two ends of the neo prene straps together and fasten with the...

Страница 10: ...ET 1313 ET 7070 offers eight preset treatment program modes The programs differ with respect to varying pulse widths and frequencies The user may choose suitable stimulation mode depend ing on their...

Страница 11: ...imes 5 10 15 20 25 30 35 40 45 50 55 and 60 minutes The therapy time will countdown on the display in 1 minute increments for the duration of your session The device turns off automatically when the t...

Страница 12: ...ly dampened sponge Do not use bleach when washing the wrap Do not dry clean your wrap Do not dry it over anything hot e g a radiator as it con tains plastic parts Ensure the wrap is completely dry bef...

Страница 13: ...t securely fastened Troubleshooting Always check the unit and accessories before use to prevent damage and defects these are some of the simple checks 1 Make sure the battery has sufficient charge and...

Страница 14: ...escription of Symbols i There are a number of technical symbols on your iReliev unit explained as follows Program Max Phase duration Rate Function mode Wave form Type P1 80mA 260uS 15Hz Constant A P5...

Страница 15: ...nufacturer s declaration electromagnetic emissions The ET 9090 ET 1313 ET 7070 is intended for use in the electromagnetic environment specified below The customer or the user of the ET 9090 ET 1313 ET...

Страница 16: ...r 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5s Not applicable IEC 60601 test level 6 kV contact 8 kV air 2 kV for power supply lines 1 kV line s and neutral 5 UT 95 dip in UT for...

Страница 17: ...the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m The ET 9090 ET 1313 ET 7070 is intended for use in the electromagnetic environ ment specified below The customer or the...

Страница 18: ...rated at a maximum output power not listed above the recom mended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the m...

Страница 19: ...of fitness for a particular purpose Warrantor undertakes no responsibility for the quality of the product except as otherwise provided herein Warrantor assumes no respon sibility that the product wil...

Страница 20: ...ExcelHealth Inc www iReliev com 1603 Hart St Southlake TX 76092 1 855 723 2582...

Отзывы: