3-29
3.6.10 3860+ Remote Radio Setup
If during repair the IVPAB015 Remote Radio assembly was replaced,
the radio will need to be programed before use. Preform the following:
3.6.10.1 Remote Radio Programing:
From Service Mode A
Press the (F3) soft key then the (F6) soft key
“Load Defaults”, note Command, Sent, and Rec. setting will scroll while
defaults are loaded. Once Scrolling stops, press (F5) soft key “Store”
and verify Rec.: and Sent: display “m>”. Press CANCEL key. Power Off
(O)
key approximately 2 +/- 0.75 second and Power Off the pump.
3.6.10.2 Flow preventer Switch Verification:
Press Power On (I) key, select New Patient and verify normal Pump dis-
play.Select channel B with the F6 Key. Open Channel B door, press
black pinch clamp in on IV Set Flow preventer assembly, verify Check
Door B flashes at top of screen and pulsing alarm sounds within 2
seconds. Note: Flow switch must be On in service menu.
3.6.10.3 Flow Stop Pull Test:
Open door verify black pinch clamp has been fully retracted (Flow stop
is closed) and Flow preventer housing is retained in the gray/violet flow
stop carrier (IVP1047/AS120).
Note: If pull test fails (black flow preventer not fully retracted) on a
new 1000 series IV set, the IVP1068 Door Clamp Clip is worn and
should be replaced. Ref: AS112A drawing in Section 5, Assemblies and
Drawings.
3.6.10.4 Green Run and Red Alarm LED Verification:
Load a fluid filled set into the pump and close the door. Press Power On
(I) key, select New Patient and verify normal Pump display. With the
Rate 1.0 highlighted, press Enter again to Select a VTBI of .1. Verify
Pump Displays “Unload Set”, open the door,remove Fluid filled tubing
from Bubble detector at bottom of Pump. Verify Display message
changes to “Close Door”. Close Door and verify “Close Door Message is
removed from top of screen and “Push Start A” is flashing at bottom of
screen. Press the start/stop key. Verify Green (run LED) begins
flashing. Open the Door and verify Red alarm Led flashes and Speaker
alarm sound. Press Alarm Cancel.
3.6.10.5 Sound Adjustment Verification:
Press the Menu Key then press (F3) to select Alarm Volume menu
item. With the Up & Down arrow keys verify the alarm volume can be
raised and lowered and saved to the appropriate setting.
3.6.10.6 3860+3860+ Remote Radio Verification
Connect 106db attenuator cable from Pump Antenna Connector to Test
Remote Antenna Connector or install Antennas on both both Pump and
Remote.
3.6.10.7
Press Power (I) on Pump. If New Patient and Same Patient
are displayed, press the (F1) soft key to select New Patient to bring up
the normal Channel A display. From the normal Channel A display,
press MENU key, press (F6) soft key to select Next Menu, press (F5)
soft key to select Set Comm. Channel, then press (F1) soft key to select
Содержание MRidium 3860
Страница 2: ......
Страница 4: ...1 4...
Страница 8: ...iv...
Страница 34: ...1 26...
Страница 50: ...2 16...
Страница 71: ...3 21 Figure 3 2...
Страница 88: ...3 38 Figure B Too Much Taper Raw Curve 3860 61 Figure C Not Enough Taper Raw Curve 3860 61...
Страница 95: ...3 45 Return the Stop Cock to the Test Position...
Страница 106: ...4 2 Figure 4 1...