3-5
pressure gauge remains at 5.0, press the F1 Key
(top left white key) to enter the Calibrate Pump
screen, then press the F5 key (Span Cal). Press the
Cancel key to return to the Service Mode display
and verify P1 and P2 are now 5 +/- 0.5.
3.2.4.2.4 Zero Verification after Adjustment
Open pressure gauge hand bulb and verify P1 and
P2 now read 0 +/- 0.2. Repeat Tare Offset and
Pressure Adjust until P1 and P2 reads displays 5 +/-
0.5 with 5 PSI applied and are 0 +/-0.5 with hand
bulb open (0 pressure applied).
Figure 3-1.
Pressure test set made using a portion of the IVP105
0 to 15 PSI pressure guage with hand bulb pump & T fitting