English
- 18 -
For both versions: When the TDS content of the pure water
produced exceeds 1-2% of the water supply TDS content,
measured using the TDS meter supplied
- Replace the 5 micron sediment filter.
- Replace the carbon block filter.
- Replace the DI resin cartridge.
STORAGE
The following proceedures should be followed to protect
the unit from bacteria fouling and damage from freezing.
See "preliminary operations" in this manual for prefilters and
membrane removal and reinstallation instructions.
Short term storage:
2-4 weeks of non use.
1. Remove both prefilter housings and dump out the water.
2. Remove the DI resin cartridge housing and dump out the
water.
3. Remove top cap from RO membrane pressure vessel.
4. Open flush valve.
5. Lay unit down and allow all water to drain out for several
minutes.
6. Reinstall top cap on pressure vessel. Do not allow membrane
to dry out.
7. Reinstall prefilters and DI resin cartridge. Do not allow DI
resin cartridge to dry out.
Long term storage (winter storage):
More than 4 weeks of
non use.
1. Remove both prefilter housings and dump out the water.
Reinstall with filters.
2. Remove the DI resin cartridge housing and dump out the
water. Reinstall housing without DI cartridge.
3. Wrap cartridge in plastic wrap and double seal it in a plastic
bags.
4. Remove top cap from RO membrane pressure vessel.
5. Remove RO membrane. Use a pliars and grip one of the
black spokes on the top of the membrane and lift. Do not grab
the small white tube.
6. Wrap membrane in plastic wrap and double seal it in a
plastic bag.
7. Reinstall cap on pressure vessel.
8. IMPORTANT NOTE: Protect from freezing. Store indoors
during winter season or whenever freezing temperatures
are possible.
9. Do not allow RO membrane of DI resin cartridge to dry out
or freeze. Damage will occur.
SYSTEM PERFORMANCE EVALUATION
In order to properly evaluate and troubleshoot the system, you
will need to perform a complete set of tests on the unit. These
tests will help identify the cause of the problem. To run these
tests, you will need the following items:
- Watch or clock with a sweep second hand.
- 3 gallon (12l) bucket.
- Volume measurement system. This could be volume gradua
-
tion marks inside the bucket or a measuring cup/container.
Flow testing procedure:
- With water running from hose being tested, direct flow into
bucket for exactly 1 minute while timing with watch.
- Measure volume of water and record the gallons per minute
Use only original attachments which offer optimum character
-
istics of quality and safety. Failure to use original attachments
absolves the manufacturer from all liability.
Stopping the Equipment without Pump Unit:
1. Open the flush valve to allow water to flush out the RO
membrane for 1 minute.
2. Turn supply water off.
3. pure water ball valve to relieve pressure in the line. Close
valve.
4. Close flush valve after pressure has bled off.
Stopping the Equipment with Pump Unit:
1. Open the flush valve to allow water to flush out the RO
membrane for 1 minute.
2. Turn motor switch OFF by pressing the push button.
3. Turn supply water off.
4. Open pure water ball valve to relieve pressure in the line.
Close valve.
5. Close flush valve after pressure has bled off.
Warning
Turn the appliance on and off by means of the main switch only
and never by plugging in or unplugging the mains plug.
Never unplug by pulling the power cable.
IMPORTANT.
As with all technical appliances, your equipment should always
be used according to the instructions.
MAINTENANCE
Only maintenance authorised in this booklet may be carried out
by the user. All other operations are prohibited.
Warning
Before carrying out maintenance of any kind, unplug the appli
-
ance from the mains.
The appliance can only be disconnected from the mains by
unplugging it.
When maintenance has been completed, before reconnecting
the appliance to the mains, check that the panels have been
replaced correctly and fixed using the screws provided.
Cleaning the appliance.
Do not use direct jets of water to clean the appliance. Use a
damp cloth only.
Clean all accessories thoroughly each time they are used.
Daily maintenance:
After the last use of the day, open the drain valve for 1 minute
to flush the concentrated water from the membrane. This will
increase the life of the membrane.
PREFILTERS - DI RESIN CARTRIDGE
Every 5000 gallons (19.000 litres) Equipment without Pump
Unit;
Every 6000 gallons (22.800 litres) Equipment with Pump Unit
of feed water run through the machine as indicated by the
totallizing flow meter;
!
!
Содержание HighPure HP B
Страница 2: ...2 Italiano Pag 6 English Pag 14 Fran ais Pag 22 Deutsch Pag 30 Espa ol Pag 38 Portugu s Pag 46 Pag 54...
Страница 54: ...54 IEC 60364 1 1 5 IEC 60364 1 I...
Страница 55: ...55 30 30...
Страница 56: ...56 RO 98 TDS TDS 200 ppm TDS 4 ppm 0 ppm 12 4 5 6 7 5 19000 22800 19000 22800 II 1 12 74 1 1 0 30...
Страница 58: ...58 E IEC 60364 1 1 2 3 4 2 5 6 fig 6 1 2 3 I 4 5 10 6 7 5 7 1 2 3 4 1 2 0 3 4 5...
Страница 62: ...62...