PURIST
OPERATING INSTRUCTIONS
Maintenance / Service
SUP_7.036_Rev01
68 (80)
Interval
Maintenance task
Personnel
As needed
Tighten bolts
Medical technician
from the user
Before each
use
Check device for defects
Specialized medical
personnel
Clean and disinfect the device
Specialized medical
personnel
Clean sealing edges of suction cups
Specialized medical
personnel
Check the accessories of the product for defects
Specialized medical
personnel
Check the tilt bearing for blocking
Specialized medical
personnel
Check traction boot for defects
Specialized medical
personnel
Check the functioning of the gas strut
Specialized medical
personnel
6.5.
Maintenance work
ATTENTION
Damage to parts caused by improper cleaning and
disinfection
Do
not
use the following products for cleaning and disinfection:
Products containing alcohol (e.g. hand disinfection agents)
•
Halides (e.g. fluorides, chlorides, bromides, iodides)
•
Halogen separating compounds (e.g. fluorine, chlorine, bromine,
iodine)
•
Products that scratch the surface (e.g. scouring agents, wire
brushes, steel wool)
•
Commercially available solvents (e.g. gasoline, thinners)
•
Water containing iron particles
•
Cleaning sponges containing iron
•
Products containing hydrochloric acid
Use only approved cleaning and disinfection agents. Use a soft,
lint-free cloth or a soft nylon brush for the cleaning of the product.