PURIST
OPERATING INSTRUCTIONS
General Safety Notes
SUP_7.036_Rev01
19 (80)
WARNING
Risk of contamination from the product
Properly disinfect traction boot before and after surgery.
Apply a bandage to avoid direct contact between patient and
traction boot during patient preparation and before the
traction boot is fitted to the leg to be operated on.
WARNING
Danger from electric shock
The application part of the device is classified as Type B.
The application of defibrillators and defibrillator monitors is to
be conducted as per manufacturer instructions
WARNING
Chemical hazard from disinfectants
The traction boot may contain residue of disinfectants.
Avoid direct contact with the patient.
Use disposable socks.
Clean contaminated traction boots by manually washing
with mild detergent and warm water.
Check if the boot is in good condition. Replace it if necessary
CAUTION
Danger from injuries caused by the traction boot (bruises,
crushing)
Frequent use and cleaning can have a negative effect on the boot
material. In operations which last several hours, pressure sores
can occur
–
especially with patients who are slimly built or have
sensitive skin.
Check if the boot is in good condition. Replace it if necessary.
Make sure of sufficient cushioning and correct fixture of the
foot in the boot. Ensure to pad the foot correctly to minimize
foot rotation and movement in the boot.