TIGER PC Software (continued)
Calibration
Use this area to define your Custom Calibration parameters.
First connect TIGER to your PC as detailed above under „Connecting your TIGER to a PC‟.
If no gas is shown in the drop down box,
browse your system to find the gas table
for your instrument. Go to the location
where the TIGER PC software files are
saved. Follow the path:
IonScience/Tiger/software/instruments/
serial no./gastable
Select the appropriate instrument
number and open the gas table.
Select the calibration gas using the drop-
down box.
TIGER PC offers a two-point calibration
(zero + span 1) or three-point (zero + span 1 + span 2) calibration. Enter the SPAN 1
concentration. For two-point calibration ensure that the box is un-ticked. For three-point
calibration tick the box and enter the SPAN 2 concentration. Send this information to your
TIGER. The calibration procedure is detailed under Maintenance on page 33 of this manual.
Startup Text
Enter the text you wish to appear on the start-up screen of your TIGER.
Send to Instrument
When you have configured your instrument or completed your changes, send them to your
TIGER by clicking the „Send to Instrument‟ box. If the message „There was a problem sending
to the inst
rument‟ appears, click OK and return to the Home screen. Repeat the procedure
under „Connecting your TIGER to a PC‟ above. If the „Summary‟ screen for your TIGER
appears, visit the configuration screen again. If not, repeat the entire procedure. If your PC still
fails to read or write to your TIGER seek advice from your distributor or from Ion Science.