PhoCheck TIGER 

Ion Science Ltd 

 

Page 15 of 43

 

 

Using your TIGER (continued) 

 

 

  HEALTH & SAFETY readings

   

Press  the  Health  &  Safety  (H&S)  soft  key  to 

display  the  H&S  symbol  with  a  question  mark  to  ask  if  the  user  wishes  to  proceed: 

    Press  ENTER  to  start  the  calculation.    The  STEL  (Short  Term  Exposure 

Level)  and  TWA  (Time  Weighted  Average)  values  are  shown.    The  fixed  LCD  H&S  icon  will 
continually  flash  while  calculating.    Pressing  the  ENTER  key  while  the  values  are  displayed 
presents the  user  with  a  screen  asking  if  the  calculation  should  stop.    Press  ENTER  again  to 
stop the calculation.  Press Esc to return to the main screen. 
 

  

GAS SELECTION  

Press  the  Gas  soft  key  to  display  an  alphabetical  letter  which 

can  be  changed  using  the  Up  and  Down  keys:   

  

Press  the  ENTER  key  to 

present  the  gases  beginning  with  that  letter.    Using  the  Up  and  Down  keys  scroll  through  the 
gases starting with that letter to find the subject gas.  Press enter to select the chosen gas.  If a 
tick appears, press Enter again and TIGER will use the relevant data for the chosen gas from 
the  gas  table  for  alarms  and  response  factor  etc.    Should  the  selected  gas  have  no  numeric 

response  factor  then  a  warning  screen

 

 

will  appear.    The  message  can  be 

cleared by pressing the ENTER key and the gas will still be selected.  An alternative lamp type 
may then be selected to match the chosen gas.  Should an incompatible gas/lamp combination 
be selected the instrument will show the warning screen again. 
 

  PEAK HOLD 

Press  the  Peak  Hold  soft  key  to  present  the  screen   

  

When  'Peak  hold'  is  selected  the  sub  display  will  appear  and  this  will  continue  to  display  the 
peak  reading  until  peak  hold  is  no  longer  required.    To  deselect  the  peak  hold  function  press 
Esc. 
 

  AVERAGE 

Press  the  Average  soft  key  to  display  a  rolling  average  over  10 

seconds.    A  tick  appears  when  time  is  up.    Pressing  the  'Average  soft  key'  again  resets  the 
calculation  even  if there  is  an  average  already  running.    Press  ESCAPE  to return  to  the  main 
screen. 
 

  OPTIONS

 

Press  the  Options  soft  key  to  give  access  to  various 

adjustable  features,  selected  with  the  UP  and  DOWN  keys.    Press  ENTER  to 
confirm selection. 
 

  BACKLIGHT 

Four options are presented: 

„Permanently Off‟, „Permanently On‟, „On in 

low ambient light

‟ and „Timed‟.  (Set the time in TIGER PC - see page 21).  Select the preferred 

option with the UP and DOWN keys and confirm with ENTER. 
 

  SOUND 

 

The  sound  options  are  presented  as  set  up  in  Tiger  PC.    Use  the 

DOWN  arrow  to  select  either  keypad  beep,  alarm  sound,  crescendo  or  percentage  volume.  
Press  ENTER  to  turn  selection  on  or  off.    For  percentage  volume,  press  ENTER  again  to 
confirm selection.  The frame will flash.  Use UP and DOWN arrows to change the percentage 
volume and press ENTER to confirm the chosen value.  Press Esc twice to return to the main 
screen. 
 

  CALIBRATION

 

 

Only two options are presented:  Factory calibration and Custom 

calibration.  Select the preferred option with the UP and DOWN keys and confirm with ENTER.  
Whichever  calibration  standard  is  selected,  the  TIGER  will  operate  against  that  calibration.  
Readings taken against unsatisfactory calibrations may be unreliable. 

 

 

 

Содержание PhoCheck TIGER

Страница 1: ...Part Number 861265 PhoCheck TIGER Instrument User Manual V1 9 Register your instrument online to receive your Extended 5 year Warranty See page 41 for details...

Страница 2: ...PhoCheck TIGER Ion Science Ltd Page 2 of 43...

Страница 3: ...n a Non Hazardous environment only BATTERY REPLACEMENT Never replace primary Alkaline battery cells while in a potentially explosive or hazardous location Use only Duracell Procell Alkaline batteries...

Страница 4: ...MC requirements Group 1 Class B equipment emissions section only EN61326 1 2006 Electrical equipment for measurement control and laboratory use EMC requirements Industrial location immunity immunity s...

Страница 5: ...etails 7 Introduction to PhoCheck TIGER 8 Getting Started 9 Understanding the Keypad 10 Understanding the Display Screen 11 13 Status Icons 11 12 Main Screen area 12 Soft Key areas 13 Using your TIGER...

Страница 6: ...he Firmware Screen 25 Downloading data logged readings 26 Health and Safety screen 27 Snapshots 27 Software disclaimers 28 Batteries 29 31 Replacement of rechargeable battery pack 30 Replacement of no...

Страница 7: ...om the users own desk top Additional features can be added without the need to return the instrument to a service centre Like it s predecessor PhoCheck the TIGER also has an intuitive graphical interf...

Страница 8: ...chargeable battery fully charged However prolonged periods of storage may result in the battery pack discharging We recommend charging the instrument for seven hours before use See the battery section...

Страница 9: ...key A single press is used as a switching operation A continuous press is used to adjust numbers or change gas selections by automatically rolling Keypad function descriptions Soft keys A B rely on g...

Страница 10: ...n an alarm condition and when health and safety readings are being collected Peak Hold When peak hold is switched on the icon appears Memory status This icon only appears if the instrument has the dat...

Страница 11: ...level has no line Alarm bell The icon consists of 2 parts the Bell and the sound bars When the instrument reaches a Low alarm the Bell and one sound bar will flash when a High alarm is reached the Bel...

Страница 12: ...oft key options They are selected using the UP and DOWN keys Those labelled optional will only appear if your TIGER has that functionality These icons can be reorganised using the TIGER PC software Bu...

Страница 13: ...lso enter Sleep mode during data logging when the duration between logs exceeds 2 minutes This is a power saving function which only occurs when a tick box is set in TIGER PC ZEROING Pressing the Zero...

Страница 14: ...een again PEAK HOLD Press the Peak Hold soft key to present the screen When Peak hold is selected the sub display will appear and this will continue to display the peak reading until peak hold is no l...

Страница 15: ...setting must never be greater than the Higher alarm INFORMATION Pressing the Information soft key allows access to a number of other information screens Use the DOWN key to scroll through the screens...

Страница 16: ...the memory stick is plugged into a USB socket the flowing screen Below left should appear If it doesn t view the content of the memory stick and double click the file named ion_cd_Tiger exe On the Ti...

Страница 17: ...e additional tasks screen below tick the box and click Next to create a desktop icon On the ready to install screen below click Install On the install screen below Click OK Click Next on the Device Dr...

Страница 18: ...d open TIGER PC The Home page should appear 2 Click on Instrument s If your TIGER has previously been connected to your PC your instruments IRN Internal Reference Number will appear Note If while usin...

Страница 19: ...d Your instrument number will appear under instrument s if it was not there already and the Instrument Summary page will appear This page gives the current status of your TIGER If the snapshot screen...

Страница 20: ...three icons represent key press alarm and crescendo Crescendo increases the sound as the upper alarm is approached Sound on each of these may be enabled or disabled by ticking or un ticking the box S...

Страница 21: ...IGER with the time on your computer Lamp Enables the selection of different lamp types you may have purchased to use with your instrument Ensure that the lamp selected is the same as that fitted in yo...

Страница 22: ...concentration For two point calibration ensure that the box is un ticked For three point calibration tick the box and enter the SPAN 2 concentration Send this information to your TIGER The calibration...

Страница 23: ...wish to add new gases to the table select the Menu tab and from the drop down menu select Allow Insert Delete This will add an additional line at the bottom of the Gas table which allows the user to...

Страница 24: ...mal Survey mode is not in an alarm condition and that no datalogging or Health Safety readings are being collected Check that the battery is sufficiently charged at least two bars Click on Check for U...

Страница 25: ...datalog screen and a list of the datalog sessions is presented Expand the datalog folder and select the session of interest Details of the data collected during that session are presented in numerica...

Страница 26: ...at your instrument is fully booted and connected to your PC as detailed on page 19 Ensure that your instrument is in normal Survey mode is not in an alarm condition and that no datalogging or Health S...

Страница 27: ...om defects in materials and workmanship under normal use and service for a period of ninety 90 days from the date of delivery as evidenced by a copy of the Receipt This limitation will apply where all...

Страница 28: ...need to remove the black rubber Boot from the TIGER during charging During charging the charger will display an orange light A green light indicates that charging is complete Discharged The battery ic...

Страница 29: ...ost easily achieved by starting at the front end of the instrument step 4 below c Release the clip at the rear end of the instrument and lift the battery pack away from the body of the instrument coin...

Страница 30: ...ore replacing the holder g Replace the pack as described above WARNING Fitting batteries or connecting the clip with the wrong polarity may result in damage to the instrument WARNING Non rechargeable...

Страница 31: ...ucted on page 29 of this manual ensuring that all connections are sound and the indicator lights on the charger are in order If the battery fails to charge fit another battery pack if available If usi...

Страница 32: ...ncentration from 10ppm You will need a cylinder of the selected gas at each of the chosen concentrations Each cylinder should be regulated for a flow rate above 250ml min Have the cylinder s of gas re...

Страница 33: ...in the transparent filter housing when the probe is removed The probe can be removed for cleaning or replacement by unscrewing it from the transparent filter housing Refit the probe using fingers onl...

Страница 34: ...fragile Handle with great care Never touch the window and do not drop To remove the Mini PID sensor for cleaning or lamp replacement first ensure that the TIGER is turned off and that you are in a cl...

Страница 35: ...as a very fine power CAS Number 1344 28 1 A full material safety data sheet MSDS is available on request from Ion science ltd The key issues are listed below Always replace the lid after using the cl...

Страница 36: ...PhoCheck TIGER Ion Science Ltd Page 37 of 43 TIGER Parts TIGER Main Assembly Front End Filter Assembly...

Страница 37: ...PID stack 861205 Protective Removable rubber boot 861230 USB Cable Angled B Connector 861413 5m Extension Hose 861414 10m Extension Hose 861415 Diluter A 31063 PID lamp cleaning kit A 830206 Decontami...

Страница 38: ...PhoCheck TIGER Ion Science Ltd Page 39 of 43 Accessories Content A 861267 Phocheck Tiger Accessory Box 846216 Removal Tool Min PID Stack A 31057 Carbon Filter Assembly Sm Blue 861221 PTFE Filter Disc...

Страница 39: ...PhoCheck TIGER Ion Science Ltd Page 40 of 43 861235 Lanyard A 31063 PID Lamp Cleaning Kit A 861229 Carbon Filter Adaptor 861476 Calibration Adaptor...

Страница 40: ...cell Procell MN1500 12 hours Data log Including date time 120 000 Alarm visual Flashing Red Amber LED Alarm audible 95 dBA 30 cm Flow Rate 220 ml min in Ambient conditions Temperature Operating 20 to...

Страница 41: ...number on page 4 17 06 2010 V0 0 63 V 1 0 0 31 1 7 Instrument Warranty Information added to page 41 Accessory images added to page 39 40 Images added on pages 33 and 34 21 07 10 V0 0 63 V 1 0 0 31 1 8...

Страница 42: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Отзывы: