PhoCheck TIGER
Ion Science Ltd
Page 16 of 43
Using your TIGER (continued)
Factory Calibration
Not for operator use.
Return to Ion Science Ltd or your distributor for calibration. (see also page 33)
Custom Calibration
see under Maintenance on page 33.
ALARMS
Upper
and Lower
alarm settings are displayed. Select the
subject alarm using the UP and DOWN keys and press ENTER. The arrow(s) on the chosen
alarm will flash. Adjust the level using the UP and DOWN keys and press ENTER. Repeat,
if necessary for the other alarm. Press Esc twice to return to the main screen.
NOTE: The Lower alarm setting must never be greater than the Higher alarm.
INFORMATION
Pressing the Information soft key allows access to a number of
other information screens. Use the DOWN key to scroll through the screens. Press ESCAPE to
return to the main screen:
First Screen
Gas selected
Upper alarm
Lower alarm
Pump flow
Second screen
Lamp selected
Date of Factory calibration
Date of Custom calibration
Date and time
Third Screen
SPAN 1
(gas concentration set in TIGER PC)
SPAN 2
(gas concentration set in TIGER PC)
PID detector
A/D reading
Fourth Screen
Temperature
(optional upgrade module)
Barometric pressure
(optional upgrade module)
Relative humidity
(optional upgrade module)
Battery type and status
Fifth Screen
STEL
TWA
IRN number (Internal Reference Number)
Firmware version
Bootloader version