44
FLASH
CODE
FAULT
EFFECT
Comments
7
Error in Shark
remote
Stops driving
•
Check remote bus cable and all
connecting plugs.
•
Replace remote.
8
Error in Shark
power module
Stops driving
•
Check all cables and connecting plugs in
the Shark system.
•
Replace power module
9
Communication
error in Shark
system
Stops driving
•
Check all cables and connecting plugs in
the Shark system.
•
Replace remote.
10
Unknown error
Varying
•
Check all cables and connecting plugs.
•
Talk to your dealer.
11
Incompatible
remote
Stops driving
•
The wrong remote type is connected.
Ensure that the power module type code
matches the remote type code.