3.
To properly tighten caster journal system and guard against
flutter, perform the following check:
a.
Tip the wheelchair forward, toward the floor.
b.
Pivot both forks
D
and casters
E
to top of their arc
simultaneously.
c.
Let casters drop to bottom of arc (wheels should swing once
to one-side, then immediately rest in a straight downward
position).
d.
Adjust locknuts
F
according to freedom of caster swing.
4.
Test wheelchair for maneuverability.
5.
Readjust locknuts if necessary, and repeat STEPS 1-3 until
correct.
6.
Snap headtube cover into the caster headtube.
7.
Reinstall retaining screws.
8.14
Removing/Installing the Footboard or
Adjusting the Footboard Height
WARNING!
–
LIMITED CLEARANCE BETWEEN
FOOTBOARD AND CASTER
- The user’s feet
must remain on the footboard while operating the
wheelchair. If the user’s feet are allowed to rest off
the side of the footboard they may come in contact
with the caster possibly resulting in injury.
Removing the Footboard
1.
Remove the socket head screw
A
, three washers
B
, spacer
C
and locknut
D
that secures the footboard
E
to the footboard
support
F
.
2.
Remove the footboard from the footboard support.
70