Invacare® TDX® SP
WARNING!
Risk of Death or Serious Injury
Failure to observe these warnings can cause an electrical
short resulting in death, serious injury, or damage to
the electrical system.
–
The POSITIVE (+) RED battery cable MUST connect
to the POSITIVE (+) battery terminal(s)/post(s).
–
The NEGATIVE (-) BLACK battery cable
MUST connect to the NEGATIVE (-) battery
terminal(s)/post(s).
–
NEVER allow any of your tools and/or battery cable(s)
to contact BOTH battery post(s) at the same time.
An electrical short may occur and serious injury or
damage may occur.
–
Install protective caps on positive and negative battery
terminals.
–
Replace cable(s) immediately if cable(s) insulation
becomes damaged.
–
DO NOT remove fuse or mounting hardware from
POSITIVE (+) red battery cable mounting screw.
1.
Remove the batteries from the wheelchair. Refer to 8.11
Removing/Installing the Batteries From/Into the Wheelchair,
page 65.
2.
Cut the tie-wraps
A
that secure the battery terminal covers
B
to the battery terminals.
3.
Slide the RED battery terminal cover back on the red battery
cable to expose the POSITIVE battery terminal
C
.
4.
Slide the BLACK battery terminal cover back on the Black
battery cable to expose NEGATIVE battery terminal
D
.
5.
Remove the locknut
E
that secures the bracket
F
of the
POSITIVE battery cable to the POSITIVE (+) battery post of the
battery.
68
1143190-N-03