16 SIEVE BED MANIFOLD UNIT
Part No 1176122
57
Invacare® Solo2™ Portable Oxygen Concentrator
14. Plug the fittings back into the bottom of the Sieve Beds for both beds.
15. Reinstall the Black Inlet Barrier and secure with 2 ty-wraps.
16. Make sure the open ends of the Heat Exchanger are plugged until ready to install into the Solo2.
17. Install the Sieve Beds into the unit and hook up wires and tubing in reverse order.
FIGURE
1
Sieve Bed Tubes attached to Heat Exchangers
Manifold Power
Connections
Sieve Bed Manifold Combo
Heat Exchanger
Valve wire pin out from Manifold
(Numbers located on board)
V1 - 4-way
V4 - Flow Sensor
V2 - Conserver
V5 - P.E.
V3 - O2 Sensor