10 REPLACING THE PATIENT OUTLET FILTER AND FILTER CONNECTOR
Part No 1176122
39
Invacare® Solo2™ Portable Oxygen Concentrator
10.2 Replacing Bulkhead Fitting
•
Required tools: 1-1/8” socket, Torx® T15 screwdriver, 1/8” Allen
wrench
1.
Remove the Indication Labels on either side of the control panel
label and save for reinstall (Detail "A").
2.
Remove the 4 screws that secure the control panel to the Solo2
shroud (Detail "B").
3.
Remove the 2 screws that secure the handle to the Solo2 unit
(Detail "B").
4.
Remove the Patient Outlet HEPA filter.
5.
Disconnect the Tubing from the Patient Outlet Fitting (Detail "D").
6.
Take a 1 1/8" notched socket and place over and loosen the nut on
the Outlet Fitting.
7.
Reassemble the unit in reverse order.
Indication Labels
DETAIL “A”
Control Panel Screws
Control Panel Screws
DETAIL “B”
DETAIL “C”
DETAIL “D”
DETAIL “E”
Patient Outlet Fitting
Patient Outlet Fitting
Nut