Invacare® Solo2™ Portable Oxygen Concentrator
5 Checking O
2
Purity
1.
Turn the unit On by pressing the On/Off button.
2.
Adjust flow to 3.0 LPM continous (refer to Solo
2
User Manual), allow to run for a minimum of 8 minutes.
3.
Connect a standard hand held oxygen analyzer to the outlet port.
4.
Follow the directions provided by the analyzer manufacturer.
5.
Repeat STEPS 2-4 for continuous flow settings 2.5, 2.0, 1.5, 1.0 and 0.5.
Oxygen purity should be checked every 3 years or 26,280 hours, whichever comes first.