SECTION 6—PREVENTIVE MAINTENANCE
Perfecto
2
™
26
FIGURE 6.2
Replace the Outlet HEPA Filter
Replacing the Compressor Inlet HEPA Filter
NOTE: For this procedure, refer to FIGURE 6.3 on page 27.
NOTE: Perform this procedure during preventive maintenance or between patients depending
upon the environment the concentrator is used in.
1. Unplug the unit.
2. Remove the filter access panel by pressing down on center tab and pulling panel out.
3. Grasp the compressor inlet HEPA filter, pull outward and up until filter is dislodged
from rubber base.
4. Discard the existing compressor inlet HEPA filter.
5. Install new compressor inlet filter by inserting the filter into the rubber base.
6. Push the filter down until the rubber base touches the edge of the filter.
7. Reinstall the filter access panel.
Tubing
Outlet HEPA Filter
Tubing