Orion
3 Hygiene
3.1 Handling of returned used products
When reconditioning or repairing returned mobility devices:
•
Take precautions for yourself and the product.
•
Use protection equipment as specified locally.
Before transport (according to Biological Agents Ordinance)
Treat product according to following process steps:
Process step
Component
Application
Conditioning technique
Work station
Manual cleaning
Surface of used device
Before repair or
reconditioning
Use saturated towel
to apply cleaning
detergent “Nücosept
special” and remove
residues after 5 min
impact.
Cleaning and
disinfection
Disinfection
Surface of used device
Before repair or
reconditioning
Use saturated
disinfectant wipes
and clean* the device
surface.
Cleaning and
disinfection
*Invacare uses detergent "Nücosept special" 1.5% in water ml/ml
Disinfection tools
•
One way wipes (fleece)
•
Brushes to clean difficult to access areas
Further information
For more information see Invacare RAN procedure and Invacare Repair and Hygiene guidelines, or contact your Invacare
Service department.
6
1608032-A