VENTURE
®
HOMEFILL II COMPRESSOR
8
COMPRESSOR OPERATION CHECKLIST
Each time the HomeFill II Compressor is used to fill a cylinder, complete the
following checklist:
NOTE: All steps mentioned below are in this procedure of the instruction sheet, unless otherwise noted.
❑
Ensure the concentrator has been on for at least twenty (20) minutes. Refer to the concen-
trator Owners Manual.
❑
Perform the prefill inspection on the cylinder.
❑
Connect the cylinder to the compressor.
❑
Turn the compressor on.
❑
Examine the indicator lights on the control panel.
❑
Remove the full cylinder.
❑
Turn the compressor off.
❑
If filling another cylinder, repeat this checklist.
CYLINDER PREFILL INSPECTION (FIGURE 1)
WARNING
ONLY use cylinders that have the Invacare Venture HomeFill II connection
and the label shown in FIGURE 1.
All cylinders must be inspected before attempting to fill - otherwise, injury or
damage may occur.
THE HYDROSTATIC TESTING DATE
Aluminum cylinders must undergo testing every five (5) years.
NOTE: DO NOT fill cylinders that have not been tested in the past five (5) years. Contact your dealer for
replacement.
COMPRESSOR OPERATION
This Procedure includes the following:
Compressor Operation Checklist
Cylinder Prefill Inspection
Connecting the Cylinder to the Compressor
Turning the Compressor On
Removing the Cylinder
Setting the Flow Knob to the Prescirbed Setting
PROCEDURE 3
COMPRESSOR
OPERATION